Acta Chir Orthop Traumatol Cech. 2009; 76(5):428-434 | DOI: 10.55095/achot2009/081
Fracture of Tibial Stem in Revision Total Knee ArthroplastyCase report
- 1 Ortopedická klinika LF UP a FN Olomouc
- 2 Ústav mechaniky těles, mechatroniky a biomechaniky, Fakulta strojního inľenýrství, VUT Brno
- 3 Laboratoř mikroskopických metod LF UP v Olomouci
An unusual case of implant failure after revision total knee arthroplasty (TKA) is presented and the case report is completed with the results of an ensuing biomechanical study made to explain possible causes leading to a fracture of the tibial stem. For this we used electron microscopic and biomechanical analyses. There was no material defect found at the site of fracture. By computerized modelling of the clinical situation, however, we found out that asymmetrical progressive osteolysis of the proximal tibia could have caused significant changes in mechanical straining associated with qualitative alterations in the process of degradation and erosion of the implant, eventually leading to a fracture of the tibial stem. In this way mechanical straining at the site of fracture could have exceeded the failure threshold of titanium alloy. Stem fracture of a current tibial component design has not yet been described in the literature. The only failure so far reported concerns the stem of a femoral component, and this has also been explained by biomechanical reasons. It follows from this study that patients with progressive osteolysis around the TKA should be followed up more frequently. The conclusions of this biomechanical analysis emphasize the importance of a thorough reconstruction of bone defects in order to improve long-term survival of the implant.
Keywords: revision, total knee arthroplasty, osteolysis, failure, fracture, tibial stem
Published: October 1, 2009 Show citation
References
- BERT, J. M., McSHANE, M.: Is it necessary to cement the tibial stem in cemented total knee arthroplasty? Clin. Orthop., 356: 73-8, 1998.
Go to original source...
Go to PubMed...
- BOURNE, R. B., FINLAY, J. B.: The influence of tibial component intramedullary stems and implant-cortex contact on the strain distribution of the proximal tibia following total knee arthroplasty. An in vitro study. Clin. Orthop., 208: 95-99, 1986.
Go to original source...
- BROOKS, P. J., WALKER, P. S., SCOTT, R. D.: Tibial component fixation in deficient tibial bone stock. Clin. Orthop., 184: 302-308, 1984.
Go to original source...
- COMPLETO, A., SIMOES, J. A., FONSECA, F., OLIVEIRA, M.: The influence of different tibial stem designs in load sharing and stability at the cement-bone interface in revision TKA. Knee, 15: 227-232, 2008.
Go to original source...
Go to PubMed...
- D'LIMA, D. D., PATIL, S., STEKLOV, N., CHIEN, S., COLWELL, C. W., Jr.: In vivo knee moments and shear after total knee arthroplasty. J. Biomech., 40 (Suppl. 1): S11-17, 2007.
Go to original source...
Go to PubMed...
- FEHRING, T. K., ODUM, S., GRIFFIN, W. L., MASON, J. B., NADAUD, M.: Early failures in total knee arthroplasty. Clin. Orthop., 392: 315-318, 2001.
Go to original source...
Go to PubMed...
- FEHRING, T. K., ODUM, S., OLEKSON, C., GRIFFIN, W. L., MASON, J. B., McCOY, T. H.: Stem fixation in revision total knee arthroplasty: a comparative analysis. Clin. Orthop., 416: 217-224, 2003.
Go to original source...
Go to PubMed...
- HAKKALAMANI, S., PRASANNA, V. K., WOOD, E. V., ACHARYA, A., PARKINSON, R. W.: Tibial-stem dissociation in a modular revision total knee arthroplasty system: a comparative clinical outcome study. J. Arthroplasty, 23: 1140-1145, 2008.
Go to original source...
Go to PubMed...
- ISSACK, P. S., COTTRELL, J. M., DELGADO, S., WRIGHT, T. M., SCULCO, T. P., SU, E. P.: Failure at the taper lock of a modular stemmed femoral implant in revision knee arthroplasty. A report of two cases and a retrieval analysis. J. Bone Jt Surg., 89-A: 2271-2274, 2007.
Go to original source...
Go to PubMed...
- JAHODA, D., NYČ, O., ©IMSA, J., KUČERA, E., HANEK, P., CHRZ, P., POKORNÝ, D., TAWA, N., LANDOR, I., SOSNA, A.: Pozdní hematogenní infekce kloubních náhrad. Acta Chir. orthop. Traum. čech., 75: 88-92, 2008.
Go to original source...
- JAZRAWI, L. M., BAI, B., KUMMER, F. J., HIEBERT, R., STUCHIN, S. A.: The effect of stem modularity and mode of fixation on tibial component stability in revision total knee arthroplasty. J. Arthroplasty, 16: 759-767, 2001.
Go to original source...
Go to PubMed...
- JIRMAN, R., VAVŘÍK, P., HORÁK, Z.: Individuální 3D tibiální augmentace pro náhrady kolenního kloubu. Acta Chir. orthop. Traum. čech., 76: 60-64, 2009.
Go to original source...
- NAZARIAN, D. G., MEHTA, S., BOOTH, R. E., Jr.: A comparison of stemmed and unstemmed components in revision knee arthroplasty. Clin. Orthop., 404: 256-262, 2002.
Go to original source...
Go to PubMed...
- ROZKYDAL, Z., JANÍK, P., JANÍČEK, P., KUNOVSKÝ, R.: Revizní náhrada kolena po aseptickém uvolnění. Acta Chir. orthop. Traum. čech., 74: 5-13, 2007.
Go to original source...
- STERN, S. H., WILLS, R. D., GILBERT, J. L.: The effect of tibial stem design on component micromotion in knee arthroplasty. Clin. Orthop., 345: 44-52, 1997.
Go to original source...
- VAN LOON, C. J., KYRIAZOPOULOS, A., VERDONSCHOT, N., De WAAL MALEFIJT, M. C., HUISKES, R., BUMA, P.: The role of femoral stem extension in total knee arthroplasty. Clin. Orthop., 378: 282-289, 2000.
Go to original source...
Go to PubMed...
- VAVŘÍK, P., LANDOR, I., DENK, F.: Klinické zhodnocení pouľití keramické femorální komponenty v konstrukci náhrady kolenního kloubu. Acta Chir. orthop. Traum. čech., 75: 436-442, 2008.
Go to original source...
- VESSELY, M. B., WHALEY, A. L., HARMSEN, W. S., SCHLECK, C. D., BERRY, D. J.: The Chitranjan Ranawat Award: Long-term survivorship and failure modes of 1000 cemented condylar total knee arthroplasties. Clin. Orthop., 452: 28-34, 2006.
Go to original source...
Go to PubMed...