Acta Chir Orthop Traumatol Cech. 2010; 77(1):13-17 | DOI: 10.55095/achot2010/002
Měření polyetylenového otěru u necementované endoprotézy kyčle in vivoPůvodní práce
- Clinic of Orthopaedics, Rudolf-Elle Hospital, Department of Orthopaedics of the Friedrich-Schiller University of Jena, Eisenberg, Germany
PURPOSE OF THE STUDY:
Recent years have seen an increase in cementless total hip endoprosthesis (THP) implantations. Easy radiological measurement of wear in cups coated with X-ray dense material, such as metal or ceramics, is not applicable. An alternative technique has thus been developed to radiographically measure wear on cementless spherical implants.
MATERIAL AND METHODS:
The authors present a method in which the center of the head of the prosthesis relative to the center of the cup was defined by graphical extrapolation. Changes between the two centers on subsequent images allowed the estimation of wear. Data from 148 randomly selected patients with cemented THP were compared with that from 50 with cementless THP (head diameter always 32 mm).
RESULTS:
Within the cemented group metal heads showed significant more wear than ceramic heads (1.244 mm vs. 0.504 mm). However, calculating the wear rate per year showed no significant difference. No significant differences were observed comparing cemented and cementless group. A comparison of the own results with those of the literature revealed analogous values.
DISCUSSION:
Data of the cementless group were comparable with those obtained from conventional wear measurements on cemented hip cups thus validating the technique.
CONCLUSION:
The method offers the potential to define wear of metal cups with a polyethylene inlay or with X-ray dense components such as ceramic or metal inlays.
Klíčová slova: wear measurement; hip endoprosthesis; cementless cup; ceramic-on-polyethylene; metal-on-polyethylene; ceramic-on-ceramic
Zveřejněno: 1. únor 2010 Zobrazit citaci
ACS | AIP | APA | ASA | Harvard | Chicago | Chicago Notes | IEEE | ISO690 | MLA | NLM | Turabian | Vancouver |
Reference
- AFIFI, K. F., JACOB, H. A. C.: Verschleißmessungen bei Hüft-Totalendoprothesen mit Polyethylenpfanne (RCH-100) und hartverchromten Protasul-10 Kopf. Z. Orthop., 119: 157-162, 1981.
Přejít k původnímu zdroji...
Přejít na PubMed...
- BABISCH, J., OETTMEIER, R., SANDER, K., BLUMENTRITT, S., STILLER, K. J.: Zur Beurteilung des Polyethylenverschleißes bei der aseptischen Endoprothesenlockerung. In: Venbrocks, R., Salis-Soglio, G. (Eds.). Jahrbuch der Orthopädie. Biermann Verlag Zülpich, 1993, 183-196.
- BOS, I., WILLMANN, G.: Morphologic characteristics of periprosthetic tissues from hip prostheses with ceramic-ceramic couples: a comparative histologic investigation of 18 revision and 30 autopsy cases. Acta orthop. scand., 72: 335-342, 2001.
Přejít k původnímu zdroji...
Přejít na PubMed...
- CAMPBELL, P., SHEN, F. W., MCKELLP, H.: Biologic and tribologic considerations of alternative bearing surfaces. Clin. Orthop., 418: 98-111, 2004.
Přejít k původnímu zdroji...
Přejít na PubMed...
- CHARNLEY, J., KAMANGAR, A., LONGFIELD, M. D.: The optimum size of prosthetic heads in relation to the wear of plastic sockets in total replacement of the hip. Med. Biol. Eng., 7: 31-39, 1969.
Přejít k původnímu zdroji...
Přejít na PubMed...
- CLARKE, I. C., GUSTAFSON, A., JUNG, H., FUJISAWA, A.: Hip-simulator ranking of polyethylene wear: comparisons between ceramic heads of different sizes. Acta orthop. scand., 67: 128-132, 1996.
Přejít k původnímu zdroji...
Přejít na PubMed...
- DORR, L. D., WAN, Z., COHEN, J.: Hemispheric titanium porous coated acetabular component without screw fixation. Clin. Orthop., 351: 158-168, 1998.
Přejít k původnímu zdroji...
- DORR, L. D., WAN, Z.: Comparative results of a distal modular sleeve, circumferential coating, and stiffness relief using the Anatomic Porous Replacement II. J. Arthroplasty, 11: 419-428, 1996.
Přejít k původnímu zdroji...
Přejít na PubMed...
- GALLO, J., HAVRANEK, V., ZAPLETALOVA, J., MANDAT, D.: Měření otěru polyetylenových jamek TEP kyčelního kloubu univerzálním měřicím mikroskopem. Charakteristiky měření. Acta Chir. orthop. Traum. čech., 73: 28-33, 2006.
Přejít k původnímu zdroji...
- HRUBINA, M., BĚHOUNEK, J., SKOTÁK, M.: Střednědobé výsledky necementované jamky TEP kyčelního kloubu Ultima u primoimplantací. Acta Chir. orthop. Traum. čech., 75: 205-211, 2008.
Přejít k původnímu zdroji...
- ILCHMANN, T., MARKOVIC, L., JOSHI, A., HARDINGE, K., MURPHY, J., WINGSTRAND, H.: Migration and wear of long-term successful Charnley total hip replacements. J. Bone Jt Surg., 80-B: 377-381, 1998.
Přejít k původnímu zdroji...
- JASTY, M., GÖTZ, D. D., BRAGDON, C. R., LEE, K. R., HANSON, A. E., ELDER, J. R., HARRIS, W. H.: Wear of polyethylene acetabular components in total hip arthroplasty. An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations. J. Bone Jt Surg., 79-A: 349-358, 1997.
Přejít k původnímu zdroji...
- MANLEY, M. T., SEREKIAN, P.: Wear debris: An environmental issue in total joint replacement. Clin. Orthop., 298: 137-146, 1994.
Přejít k původnímu zdroji...
- MANNING, D. W., CHIANG, P. P., MARTELL, J. M., GALANTE, J. O., HARRIS, W. H.: In vivo comparative wear study of traditional and highly cross-linked polyethylene in total hip arthroplasty. J. Arthroplasty, 20: 880-886, 2005.
Přejít k původnímu zdroji...
Přejít na PubMed...
- MARTELL, J. M., BERDIA, S.: Determination of polyethylene wear in total hip replacements with use of digital radiographs. J. Bone Jt Surg., 79-A: 1635-1641, 1997.
Přejít k původnímu zdroji...
Přejít na PubMed...
- MARTELL, J. M., BERKSON, E., BERGER, R., JACOBS, J.: Comparison of two and three-dimensional computerized polyethylene wear analysis after total hip arthroplasty. J. Bone Jt Surg., 85-A: 1111-1117, 2003.
Přejít k původnímu zdroji...
Přejít na PubMed...
- MARTELL, J. M., LEOPOLD, S. S., LIU, X.: The effect of joint loading on acetabular wear measurement in total hip arthroplasty. J. Arthroplasty, 15: 512-518, 2000.
Přejít k původnímu zdroji...
Přejít na PubMed...
- MURRAY, D. W., RUSHTON, N.: Macrophages stimulate bone resorption when they phagocytose particles. J. Bone Jt Surg., 72-B: 988-992, 1990.
Přejít k původnímu zdroji...
Přejít na PubMed...
- PETRUŠKA, J., ĎURČANSKÝ, D., MAKAREVIČ, A., KUBOVIČOVÁ, E., PIVKO, J.: Morfologicko-funkčná charakteristika periartikulárneho tkaniva po endoprotéze bedrového kĺbu: histologické, cytochemické a elektrónmikroskopické aspekty. Acta Chir. orthop. Traum. čech., 75: 375-381, 2008.
Přejít k původnímu zdroji...
- SARMIENTO, A., LATTA, A. A.: Surgical Technique for the Prevention of Femoral Lysis in Cement Total Hip Arthroplasty. Acta Chir. orthop. Traum. čech., 76: 85-89, 2009.
Přejít k původnímu zdroji...
- SCHMALZRIED, T. P., GUTTMANN, D., GRECULA, M., AMSTUTZ, H. C.: The relationship between the design, position, and articular wear of acetabular components inserted without cement and the development of pelvic osteolysis. J. Bone Jt Surg., 76-A: 677-688, 1994.
Přejít k původnímu zdroji...
Přejít na PubMed...
- SCHMALZRIED, T. P., SHEPHERD, E. F., DOREY, F. J., JACKSON, W. O., DELA ROSA, M., FA'VAE, F., McKELLP, H. A., McCLUNG, C. D., MARTELL, J., MORELAND, J. R., AMSTUTZ, H. C.: The John Charnley Award. Wear is a function of use, not time. Clin. Orthop., 381: 36-46, 2000.
Přejít k původnímu zdroji...
Přejít na PubMed...
- SEMLITSCH, M., WILLERT, H. G.: Clinical wear behaviour of ultra-high molecular weight polyethylene cups paired with metal and ceramic ball heads in comparison with to metal-on-metal pairings of hip joint replacements. Proc. Inst. Mech. Eng., 211: 73-88, 1997.
Přejít k původnímu zdroji...
Přejít na PubMed...
- SHAVER, S. M., BROWN, T. D., HILLIS, S. L., CALLAGHAN, J. J.: Digital edge-detection measurement of polyethylene wear after total hip arthroplasty. J. Bone Jt Surg., 79-A: 690-700, 1997.
Přejít k původnímu zdroji...
Přejít na PubMed...
- WILLERT, H. G., SEMLITSCH, M., BUCHHORN, G., KRIETE, U.: Materialverschleiß und Gewebereaktion bei künstlichen Gelenken. Orthopäde, 7: 62-83, 1978.
- WILLMANN, G.: Ceramics for total hip replacement - What a surgeon should know. Orthopaedics, 21: 173-177, 1998.
Přejít k původnímu zdroji...
Přejít na PubMed...
- WROBLEWSKI, B. M., SINEY, P. D., DOWNSON, D.: Prospective clinical and joint simulator studies of a new total hip arthroplasty using alumina ceramic heads and cross-linked polyethylene cups. J. Bone Jt Surg., 78-B: 280-285, 1996.
Přejít k původnímu zdroji...
- YANG, W. E., SHIH, C. H.: Porous coated anatomic total hip arthroplasty: 5- to 10-year follow up. Changgeng Yi Xue Za Zhi, 21: 397-402, 1998.
- ZICHNER, L. P., WILLERT, H. G.: Comparison of alumina-polyethylene and metal-polyethylene in clinical trials. Clin. Orthop., 282: 86-94, 1992.
Přejít k původnímu zdroji...
- ZILLER, R., UHLICH, T., BLUMENTRITT, S., MANNER, P.: Röntgenologische Verschleißmessung an der Polyäthylenpfanne bei Hüfttotalendoprothesen. Beitr. orthop. Traum., 32: 290-296, 1985.