Acta Chir Orthop Traumatol Cech. 2010; 77(3):186-193 | DOI: 10.55095/achot2010/039
Results Covering 20 Years Use of the Cement-free Zweymüller Alloclassic Total Endoprosthesis of the Hip JointOriginal papers
- Centre for Orthopaedic and Orthopaedic Surgery of the Landesklinikum Waldviertel Zwettl, Austria
PURPOSE OF THE STUDY:
From an initial group of 206 hips in 190 patients we carried out a retrospective clinical and radiological control of 74 hips in 72 patients covering a period of 20 years.
MATERIAL: AND METHOD
Indications in 74 operated hips evidenced 48 x (64,9 %) primary and 26 x (35,1 %) secondary coxarthrosis. The average age at the time of operation was 53 years (37 years - 68 years). All 72 patients with 74 hips received cement-free self-cutting Alloclassic screw cups of pure Titanium together with a cement-free square Alloclassic stem consisting of Titan-Aluminium-Niob alloy according to Zweymüller.
Regarding the technique of operation we used the approach after Watson-Jones in the supine position. As prophylaxis against thrombosis 40 % of the patients received derivates of heparine and 60 % had anticoagulant therapy with cumarine.
Prophylaxis against heterotopic ossifications were not carried out at this juncture in our department.
RESULTS:
After an average post-operational examination lapse of 20 years we were able to examine 72 patients (37,9 %) with 74 hips (35,9 %) clinically and radiologically.
Further classification of the 72 patients records 47 females and 25 males. The average age at the time of operation was 53 years and at the time of post-operational check-up to 74 years. For the clinical post-operational check-up (n = 74) we used the Harris Hip Score.
Further evaluation shows 12 x (16,2 %) excellent results, 26 x (35,1 %) good and 29 x (39,2 %) fair results. In 7 patients (9,5 %) we had to observe poor results because of multimorbidity, although also in these cases stability of TEP had been achieved.
The radiological post-operational check-up of the 74 stable hips (35,9 %) shows an average excentric position of the head of 1,4 mm (0 mm up until 4 mm) compared with 1 mm after 10 years.
Over an average of about 20 years we carried out an exchange of the inlay and the head because of excentric position of the head, which correlates to a rate of reoperation of 6,8 %.
DISCUSION
In 2000 and 2001 we published our 10 year results with cement-free Alloclassic screw cup and cement-free stem in 133 hips in 123 patients and compared our findings with those of other authors. After 10 years we saw only 3 % complications with cups and an overall re-operation rate of 6,7 %. The 20-year-results show in all cases stable components of both cups and stems. 5 patients (6,8 %) had to undergo re-operation with an exchange of inlay and head, whereby the components of the prosthesis themselves turned out to be stable.
An average polyethylene abrasion of 1 mm after 10 years compares with a value of abrasion of 1,4 mm after 20 years.
The 6,8 % rate of re-operation after 20 years indicates the excellent results of the use of cement-free Alloclassic total endoprosthesis system of the hip according to Zweymüller.
CONCLUSIONS:
The biocompatible qualities of modern prosthesis material lead to a quicker and optimal bony incorporation of the prosthesis components. The excellent 20-year-results show a broad indicative spectrum, component stability of the prosthesis in all patients and therefore encourage the further use of cement-free hip implants in the future.
Keywords: alloclassic - Total Endoprosthesis Systems after Zweymüller, 20-years-results
Published: June 1, 2010 Show citation
ACS | AIP | APA | ASA | Harvard | Chicago | Chicago Notes | IEEE | ISO690 | MLA | NLM | Turabian | Vancouver |
References
- BAUMGARTNER, R., MÜLLER, W., LINDENFELD, T.: Erfahrungen mit Komponenten des Alloclassic-Systems ab 1985. In: Schmitt, M. (Ed.): Die Metallpaarung <Metasul> in der Hüftendoprothetik. Bern, Stuttgart, Toronto, Verlag Hans Huber 1995, 115-124.
- BÖHLER, N., GRIEBLER, W., AUERSPERG, V.: Mittelfristige Ergebnisse des Alloclassic-Hüftsystems. In: Schmitt, M. (Ed.). Die Metallpaarung in der Hüftendoprothetik. Bern, Stuttgart, Toronto, Verlag Hans Huber 1995, 125-136.
- BÖHM, G., LINTNER, F., KLIMANN, S., HUBER, M.: Histomorphologic and Morphometric Evaluation of Cementless Implanted Hip Endoprostheses. Int. J. Surg. Pathol. 2 (Suppl. 9): 22-25, 1995.
- BRAGDON, C., GREENE, M., GOLDVASSER, D., FREIBERG, A., MALCHAU, H.: Mid-term Clinical and Wear Results of Highly Cross-linked Polyethylene Acetabular Components. EFORT congress, Nice, Abstract F10, 2008.
- DELAUNAY, C., KAPANDJI, A. I.: Cementless primary total hip replacement. 4-8 year results with the Zweymüller-Alloclassic prosthesis. Int. Orthop., 22: 1-5, 1998.
Go to original source...
Go to PubMed...
- DELAUNAY, C., KAPANDJI, A. I.: Primary total hip arthroplasty with the Zweymüller first-generation cementless prosthesis. A 5-9 year retrospective study. J. Arthroplasty, 11: 643-652, 1996.
Go to original source...
Go to PubMed...
- DIGAS, G., KÄRRHOLM, J., THANNER, J., HERBERTS, P.: 5-year experience of highly cross-linked polyethylene in cemented and uncemented sockets. Two randomized studies using radiostereometric analysis. Acta Orthop., 78: 746-754, 2007.
Go to original source...
Go to PubMed...
- FEIGHAN, J. E., GOLDBERG, V. M., DAVY, D. et al.: The influence of surfaceblasting on the incorporation of titanium-alloy implants in a rabbit intramedullary model. J. Bone Jt Surg., 77-A: 1380-1395, 1995.
Go to original source...
Go to PubMed...
- GRÜBL, A., CHIARI, C., GRUBER, M., KAIDER, A., GOTTSAUNER-WOLF, F.: Cementless total hip arthrosplasty with a tapered, rectangular titanium stem and a threaded cup. J. Bone Jt Surg., 84-A: 425-431, 2002.
Go to original source...
Go to PubMed...
- GRÜBL, A., CSEPAN, R., DELAUNAY, C., GÖRDES, W., KAIDER, A., PARZER, R., ZENZ, P., GOTTSAUNER-WOLF, F.: 6-10-Jahres-Ergebnisse des zementfreien Alloclassic-Hüftsystems - Eine multizentrische Survival-Studie. Z. Orthop., 141: 303-308, 2003.
Go to original source...
Go to PubMed...
- KUTSCHERA, H. P., EYB, R., SCHARTELMÜLLER, T., TOMA, C., ZWEYMÜLLER, K.: Das zementfreie Zweymüller-Hüft-System. Ergebnisse einer 5-Jahres-Nachuntersuchung. Z. Orthop., 131: 513-517, 1993.
Go to original source...
Go to PubMed...
- LANDOR, I., VAVØÍK, P., JAHODA, D., POKORNÝ, D., BALLAY, R., SOSNA, A.: Dlouhodobé zku¹enosti s kombinovaným hydroxyapatitovým povrchem ARBOND v osteointegraci implantátu. Acta Chir. orthop. Traum. èech., 76: 172-178, 2009.
Go to original source...
- LINTNER, F., BÖHM, G., BRAND, G.: Bone Reactions to Hip Joint Replacements made of Titanium Alloys. Titanium Developement Association. Titanium International Conference, 1990, 672-688.
- LINTNER, F., ZWEYMÜLLER, K., BRAND G.: Reactions of Surrounding Tissue to the Cementless Hip Implant Ti-6Al-4V after an Implantation Period of several years. Autopsy studies in three cases. Arch. Orthop. Trauma Surg., 107: 357-363, 1988.
Go to original source...
Go to PubMed...
- OLYSLAEGERS, C., DEFOORT, K., SIMON, J. P., VANDENBERGHE, L.: Wear in conventional and highly cross-linked polyethylene cups. J. Arthroplasty, Vol. 23, 2008.
Go to original source...
Go to PubMed...
- PETRU©KA, J., ÏURÈANSKÝ, D., MAKAREVIÈ, A., KUBOVIÈOVÁ, E., PIVKO, J.: Morfologicko-funkèná charakteristika periartikulárneho tkaniva po endoprotéze bedrového kåbu: histologické, cytochemické a elektrónmikroskopické aspekty. Acta Chir. orthop. Traum. èech., 75: 375-381, 2008.
Go to original source...
- POKORNÝ, D., ©LOUF, M., DYBAL, J., ZOLOTAREVOVÁ, E., VESELÝ, F., JAHODA, D., VAVØÍK, P., LANDOR, I., ENTLICHER, G., SOSNA, A.: Nová metoda kvantifikace otìrových èástic UHMWPE v okolí kloubních náhrad. Acta Chir. orthop. Traum. èech., 76: 374-381, 2009.
Go to original source...
- POSPISCHILL, M., KNAHR, K.: Cementless total hip arthroplasty using a threaded cup and a rectangular tapered stem. J. Bone Jt Surg., 87-B: 1210-15, 2005.
Go to original source...
Go to PubMed...
- RANDELLI, G., ROMANO, C., VISENTIN, O.: Longterm results of Alloclassic-Zweymüller hip prosthesis. J. Bone Jt Surg., 79-B (Suppl. 2): 238-238, 1997.
- ROZKYDAL, Z., JANåÈEK, P., HAVLåÈEK, V., PAZOUREK, L.: Dlouhodobé výsledky CLS døíku u primární náhrady kyèle. Acta Chir. orthop. Traum. èech., 76: 281-287, 2009.
Go to original source...
- ROZKYDAL, Z., JANåÈEK, P., TOMÁ©, T., FLORIAN, Z.: Dlouhodobé výsledky CLS jamky u primární náhrady kyèle. Acta Chir. orthop. Traum. èech., 76: 90-97, 2009.
Go to original source...
- SEMLITSCH, M., STAUB, F., WEBER, H.: Titanium-Aluminium-Niobium Alloy, Development for Biocompatible Highstrength Surgical Implants. Biomed. Tech., 30: 334-339, 1985.
Go to original source...
Go to PubMed...
- SEMLITSCH, M.: Titanium Alloys for Hip Joint Replacement. Clinical Materials, 2: 1-13, 1987.
Go to original source...
- SOSNA, A., RADONSKÝ, T., POKORNÝ, D., VEIGL, D., HORÁK, Z., JAHODA, D.: Polyetylénová choroba. Acta Chir. orthop. Traum. èech., 70: 6-16, 2003.
- VERVEST, T. M. J. S., ANDERSON, P. G., FREEK VAN HOUT, WAPSTRA, F. H, LOUWERSE, R. T., KOETSIER J. W. A: Ten to Twelve-Year results with the Zweymüller cementless Total hip prosthesis. J. Arthoplasty, 20: 362-368, 2005.
Go to original source...
Go to PubMed...
- VERVEST, T. M. J. S., ANDERSON, P. G.: The Zweymüller cementless total hip prosthesis in patients aged 50 years and younger. Hip International, 15: 1-11, 2005.
Go to original source...
- WEISSINGER, M., HELMREICH, C.: Langfristige Resultate mit dem zementfreien Alloclassic-Schaft nach Zweymüller. Z. Orthop., 139: 200-205, 2001.
Go to original source...
Go to PubMed...
- WEISSINGER, M., HELMREICH, C.: Zehnjährige Erfahrungen mit der zementfreien Alloclassic-Pfanne nach Zweymüller. Acta Chir. orthop. Traum. èech., 67: 157-163, 2000.
- WEISSINGER, M., KEPPEL, A., HELMREICH, C.: Retrospektive Studie über heterotope Ossifikationen mit und ohne Indometacinprophylaxe. Acta Chir. orthop. Traum. èech., 61:87-91, 1994.
- ZENZ, P., POSPISIL, C., FERTSCHAK, W., SCHWÄGERL, W.: 10 Jahre zementfreie Implantation von Hüfttotalendoprothesen unter Verwendung des Zweymüller-Schaftes. Z. Orthop., 133: 558-561, 1995.
Go to original source...
Go to PubMed...
- ZWEYMÜLLER, K., DECKNER, A., LINTNER, F., SEMLITSCH, M.: Die Weiterentwicklung des zementfreien Systems durch das SL-Schaftprogramm. Med. Orthop. Tech., 108: 10-15, 1988.
- ZWEYMÜLLER, K., SAMEK, V.: Radiologische Erkenntnisse der Titaniumpfanne. In: Zweymüller, K. (Hrsg.): 10 Jahre Zweymüller-Hüftendoprothese. Bern, Stuttgart, Toronto, Huber 1990.
- ZWEYMÜLLER, K., SEMLITSCH, M.: Concept and Material Properties of a Cementless Hip Prosthesis System with Al2O3 Ceramic Ball Heads and Wrought Ti-6Al-4V Stems. Arch. Orthop. Traumat. Surg., 100: 229-236, 1982.
Go to original source...
Go to PubMed...
- ZWEYMÜLLER, K., SEMLITSCH, M.: Weiterentwicklung des zementfreien Hüftendoprothesensystems aufgrund spezieller Indikationen. Med. Orthop. Tech., 106:11-14, 1986.