Acta Chir Orthop Traumatol Cech. 2011; 78(5):410-415 | DOI: 10.55095/achot2011/063
Sérové hladiny kobaltu u bezcementových totálních náhrad kyčelního kloubu s párováním kov-kov; výsledky dvouleté prospektivní studiePůvodní práce
- Centre for Orthopaedics and Orthopaedic Surgery of the Landesklinikum Waldviertel Zwettl, Austria
PURPOSE OF THE STUDY:
Total hip arthroplasty increases the use of alternate bearings to prevent polyethylene wear as the number of younger and more active patients has drastically risen.
We carried out a prospective randomized study, to assess and compare clinical results and radiological changes, serum-cobalt- and serum-aluminium-levels when metal-on-metal and ceramic-on-ceramic bearings are applicated.
MATERIAL AND METHODS:
After giving informed consent 80 consecutive patients were included in this prospective randomized study. They were randomly assigned to receive either a metal-on-metal or a ceramic-on-ceramic bearing in their total hip replacement. Eligible were patients with a primary coxarthrosis or an avascular necrosis of the head of femur. Of the 80 patients 54 were females and 26 males. 42 patients were randomized to a metal-on-metal bearing and 38 patients were randomized to a ceramic-on-ceramic bearing. The average patient-age was 65,8 years and the mean body mass index was 27,7 at the time of operation. Surgery was performed through a transgluteal approach in supine position under general or spinal anaesthesia. A forged conical threaded acetabular component made of titanium-aluminium-niobium alloy was used in all patients. The metal inlays and the 28 mm metal heads were made of Co-28Cr-6Mo alloy with a carbon content of 0,2%. The ceramic inlays and the 28 mm ceramic heads are Al2O3 implants. We used as femoral component a conical rectangular stem of a titanium-aluminium-niobium alloy. Cup and stem werde implanted cementfree.
Clinical data werde obtained at a follow up at a minimum of two years after implantation.
Patients were assessed with the Harris Hip Score and the University of California at Los Angeles activity scale. 72 of the 80 patients could be explored clinically and radiologically.
RESULTS:
The 2 year follow up check showed clinically and radiologically no difference between the two groups. The median Harris HipScore was above 90 points and the UCLA score was about 7 points. The medium serum-cobalt level was in the metal-on-metal group about 1.2μg/L and in the ceramic-on-ceramic group below the detection limit. The medium serum aluminium level showed values of 1.2 respectively 1.3 μg/L. The luminescencies in the metal-on-metal group were increased, but all components of the prosthesis could be regarded as stable.
DISCUSSION:
It was the goal of our prospective study to compare clinical and radiological results of hip arthroplasty in metal-on-metal and ceramic-on-ceramic bearings and assess the serum-cobalt and the serum-aluminium levels. The socalled "biocorrosion" is still a high risk element for loosening of implants because of aseptic osteolysis. A higher release of polyethylene and metal particles is triggered, which leads to a slow but continuous process of inflammation.
Apart from the debris also metal ions in a higher concentration are released, so that we could detect after some years in patients with artificial implants a higher level of metal in blood and urine. So far we did not recognize any carcinogenity or the appearance of renal disorders, or other mutagene effects in our patients. We could also not observe any fractures in ceramic implants. We are convinced that short time and intermediate reports are necessary, although final judgement can only be based on long term data.
CONCLUSION:
Our prospective randomized study showed after two years no difference clinically between the two groups of metal-on-metal and ceramic-on-ceramic bearings with total endoprostheses of the hip. Although medium serum-cobalt level in the metal-on-metal group with 1,2μ/L is a significant higher value, whereas it lies in the ceramic-on-ceramic group below the detectable limit.
Klíčová slova: total endoprosthesis of the hip, metal-on-metal bearing, serum-cobalt level, metallosis
Zveřejněno: 1. říjen 2011 Zobrazit citaci
ACS | AIP | APA | ASA | Harvard | Chicago | Chicago Notes | IEEE | ISO690 | MLA | NLM | Turabian | Vancouver |
Reference
- AMSTUTZ, H. C., LE DUFF, M. J.: Eleven years of experience with metal-on-metal hybrid hip resurfacing. J. Arthroplasty, 23: 36-43, 2008.
Přejít k původnímu zdroji...
Přejít na PubMed...
- BOEHLER, M., PLENK, H., SALZER, M.: Alumina Ceramic Bearings for Hip Endoprostheses. The Austrian Experience. Clin. Orthop., 379: 85-93, 2000.
Přejít k původnímu zdroji...
Přejít na PubMed...
- BOUTIN P.: Total arthroplasty of the hip by fritted aluminium prosthesis: experimental study and 1st clinical applications. Rev. Chir. Orthop. Reparatrice Appar Mot., 58: 230-246, 1972.
- BRODNER, W., BITZAN, P., MEISINGER, V., KAIDER, A., GOTTSAUNER-WOLF, F., KOTZ, R.: Serum cobalt levels after metal-on-metal total hip arthroplasty. J. Bone Jt Surg., 85-A: 2168-2173, 2003.
Přejít k původnímu zdroji...
Přejít na PubMed...
- BRODNER, W., GRÜBL, A., JANKOVSKY, R., MEISINGER, V., LEHR, S., GOTTSAUNER-WOLF, F.: Cup inclination and serum concentration of cobalt and chromium after metal-on-metal total hip arthroplasty. J. Arthroplasty, 19: 66-70, 2004.
Přejít k původnímu zdroji...
Přejít na PubMed...
- BROOKER, A., BOWERMAN, J., ROBINSON, R.: Ectopic ossification following total hip replacement. J. Bone Jt Surg., 55-A: 1629-1632, 1973.
Přejít k původnímu zdroji...
- CADOSCH, D., SCHLETT, C. L., GAUTSCHI, O. P., FREI, H. C., FILGUEIRA, L.: Metallionen: Wichtige Mitspieler in der aseptischen Lockerung. Z. Orthop. Unfall., 148: 393-397, 2010.
Přejít k původnímu zdroji...
Přejít na PubMed...
- CLARKE, I. C., GOOD, V., WILLIAMS, P., SCHROEDER, D., ANISSIAN, L., STARK, A., OONISHI, H., SCHULDIES, J., GUSTAFSON, G.: Ultra-low wear rates for rigid-on-rigid bearings in total hip replacements. Proc. Inst. Mech. Eng. (H), 214: 331-347, 2000.
Přejít k původnímu zdroji...
Přejít na PubMed...
- D'ANTONIO, J., CAPELLO, W., MANLEY, M., NAUGHTON, M., SUTTON, K.: Alumina Ceramic Bearings for Total Hip Arthroplasty. Clin. Orthop., 436: 146-171, 2005.
Přejít k původnímu zdroji...
- DAVIES, A. P., SOOD, A., LEWIS, A. C., NEWSON, R., LEARMONTH, I. D., CASE, C. P.: Metal-specific differences in levels of DNA damage caused by synovial fluid recovered at revision Arthroplasty. J. Bone Jt Surg., 87-B: 1439-1444, 2005.
Přejít k původnímu zdroji...
Přejít na PubMed...
- DELEE, G., CHARNLEY, J.: Radiological Demarcation of Cemented Sockets in Total Hip Replacement. Clin. Orthop., 121: 20-32, 1976.
Přejít k původnímu zdroji...
- ESSNER, A., SUTTON, K., WANG, A.: Hip simulator wear comparison of metal-on-metal, ceramic-on-ceramic and cross linked UHMWPE bearings. Wear, 259: 992-995, 2005.
Přejít k původnímu zdroji...
- FIRESTONE, D. E., CALLAGHAN, J.J., LIU, S. S., GOETZ, D. D., SULLIVAN, P. M., VITTETOE, D. A., JOHNSTON, R. C.: Total hip arthroplasty with a cemented, polished, collared femoral stem and a cementless acetabular component. A follow-up study at a minimum of ten years. J. Bone Jt Surg., 89-A: 126-132, 2007.
Přejít k původnímu zdroji...
Přejít na PubMed...
- GRÜBL, A., CHIARI, C., GIUREA, A., GRUBER, M., KAIDER, A., MARKER, M., ZEHETGRUBER, H., GOTTSAUNER-WOLF, F.: Cementless Total Hip Arthroplasty with the Rectangular Titanium Zweymüller Stem. A Concise Follow-up, at a Minimum of fifteen Years, of a Previous Report. J. Bone Jt Surg. Am., 88(10): 2210-2215, 2006.
Přejít k původnímu zdroji...
- GRÜBL, A., CHIARI, C., GRUBER, M., KAIDER, A., GOTTSAUNER-WOLF, F.: Cementless Total Hip Arthroplasty with a Tapered, Rectangular Titanium Stem and a Threaded Cup. J. Bone Jt Surg., 84-A: 425-431, 2002.
Přejít k původnímu zdroji...
Přejít na PubMed...
- GRÜBL, A., MARKER, M., BRODNER, W., GIUREA, A., HEINZE, G., MEISINGER, V., ZEHETGRUBER, H., KOTZ, R.: Long-term follow-up of metal-on-metal total hip replacement. J. Orthop. Res., 25: 841-848, 2007.
Přejít k původnímu zdroji...
Přejít na PubMed...
- GRÜBL, A., WEISSINGER, M., BRODNER, W., GLEISS, A., GIUREA, A., GRUBER, M., PÖLL, G., MEISINGER, V., GOTTSAUNER-WOLF, F., KOTZ, R.: Serum aluminium and cobalt levels after ceramic-on-ceramic and metal-on-metal total hip replacement. J. Bone Jt Surg., 88-B: 1003-1005, 2006.
Přejít k původnímu zdroji...
Přejít na PubMed...
- GRUEN, T. A., MCNEICE, G. M., AMSTUTZ, H. C.: Modes of Failure of Cemented Stem-type Femoral Components. A Radiographic Analysis of Loosening. Clin. Orthop., 141: 17-27, 1979.
Přejít k původnímu zdroji...
- HA, Y. C., KIM, S. Y., KIM, H. J., YOO, J. J., KOO, K. H.: Ceramic Liner Fracture after Cementless Alumina-on-Alumina Total Hip Arthroplasty. Clin. Orthop., 458: 106-110, 2007.
Přejít k původnímu zdroji...
Přejít na PubMed...
- HARRIS, W. H.: The problem is osteolysis. Clin. Orthop., 311: 46-53, 1995.
- HARRIS, W. H.: Traumatic Arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J. Bone Jt Surg., 51-A: 737-755, 1969.
Přejít k původnímu zdroji...
- HASEGAWA, M., SUDO, A., UCHIDA, A.: Alumina ceramic-on-ceramic total hip replacement with a layered acetabular component. J. Bone Jt Surg., 88-B: 877-882, 2006.
Přejít k původnímu zdroji...
Přejít na PubMed...
- JANÍČEK, P., SKLENSKÝ, M., KLEPÁČ, L., PELCLOVÁ, D.: Cobalt intoxikation due to head prothesis wear in total hip replacement. XIV.Národni kongres ČSOT: 19.-21.5. 2010, Ortopedie, 4: 18, 2010.
- LEARMONTH, I. D., CASE, C. P.: Metallic debris from orthopaedic implants. The Lancet, 369: 542-544, 2007.
Přejít k původnímu zdroji...
Přejít na PubMed...
- PARK, Y. S., HWANG, S. K., CHOY, W. S., KIM, Y. S., MOON, Y. W., LIM, S. J.: Ceramic Failure after Total Hip Arthroplasty with an Alumina-on-Alumina Bearing. J. Bone Jt Surg., 88-A: 780-787, 2006.
Přejít k původnímu zdroji...
Přejít na PubMed...
- POKORNÝ, D., ŠLOUF, M., DYBAL, J., ZOLOTAREVOVÁ, E., VESELÝ, F., JAHODA, D., VAVŘÍK, P., LANDOR, I., ENTLICHER, G., SOSNA, A.: Nová metoda kvantifikace otěrových částic UHMWPE v okolí kloubních náhrad. Acta Chir. orthop. Traum. čech., 76: 374-381, 2009.
Přejít k původnímu zdroji...
- PRUDHOMMEAUX, F., HAMADOUCHE, M., NEVELOS, J., DOYLE, C., MEUNIER, A., SEDEL, L.: Wear of alumina-on-alumina total hip arthroplasties at a mean 11-year follow -up. Clin. Orthop., 379: 113-122, 2000.
Přejít k původnímu zdroji...
Přejít na PubMed...
- ROTH, A., SANDER, K., LAYHER, F., BABISCH, J., VENBROCKS, R.: In Vivo Measurement of Polyethylene Wear in Cementless Total Hip Arthroplasty. Acta Chir. orthop. Traum. čech., 77: 13-17, 2010.
Přejít k původnímu zdroji...
- SIEBER, H. P., RIEKER, C. B., KOTTIG, P.: Analysis of 188 second-generation metal-on-metal retrieved hip implants. J. Bone Jt Surg., 81-B: 46-50, 1999.
Přejít k původnímu zdroji...
- SIGNORELLO, L. B., YE, W., FRYZEK, J. P., BLOT, W. J., LIPWORTH, L., MCLAUGHLIN, J. K., NYRÉN, O.: A nationwide follow-up study of autoimmune and connective tissue disease among hip and knee implant patients. J. long-term eff. med. implants, 12: 255-262, 2002.
Přejít k původnímu zdroji...
- TONI, A., TRAINA, F., STEA, S., SUDANESE, A., VISENTIN, M., BORDINI, B., SQUARZONI, S.: Early Diagnosis of Ceramic Liner Fracture. Guidelines Based on a twelve-Year Clinical Experience. J. Bone Joint Surg., 88-A: 55-63, 2006.
Přejít k původnímu zdroji...
- TRICLOT, P.: Total hip replacement with metal on metal bearing system with a high content in carbon. Results at minimum 12 years of a consecutive series. Rev. Chir. Orthop., 95: 85-91, 2009.
- VISURI, T., PUKKALA, E., PAAVOLAINEN, P., PULKKINEN, P., RISKA, E. B.: Cancer risk after metal on metal and polyethylene on metal total hip Arthroplasty. Clin. Orthop., 329S: 280-289, 1996.
Přejít k původnímu zdroji...
Přejít na PubMed...
- WALTER, W. L., O'TOOLE, G. C., WALTER, W. K., ELLIS, A., ZICAT, B. A.: Squeaking in Ceramic-on-Ceramic Hips. The Importance of Acetabular Component Orientation. J. Arthroplasty, 22: 496-503, 2007.
Přejít k původnímu zdroji...
Přejít na PubMed...
- WEBER, B. G.: Metal-metal total prosthesis of the hip joint: back to the future. Z. Orthop. Ihre Grenzgeb., 130: 306-309, 1992.
Přejít k původnímu zdroji...
Přejít na PubMed...
- WEISSINGER, M., HELMREICH, C.: Langfristige Resultate mit dem zementfreien Alloclassic-Schaft nach Zweymüller. Z. Orthop. Ihre Grenzgeb., 139: 200-205, 2001.
Přejít k původnímu zdroji...
Přejít na PubMed...
- WEISSINGER, M., HELMREICH, C., PÖLL, G.: Results Covering 20 Years Use of the Cement-free Zweymüller Alloclassic Total Endoprosthesis of the Hip Joint. Acta Chir. orthop. Traum. čech., 77: 186-193, 2010.
Přejít k původnímu zdroji...
- WILLERT, H. G., BUCHHORN, G. H., FAYYAZI, A., FLURY, R., WINDLER, M., KÖSTER, G., LOHMANN, C. H.: Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study. J. Bone Jt Surg., 87-A: 28-36, 2005.
Přejít k původnímu zdroji...
Přejít na PubMed...