Acta Chir Orthop Traumatol Cech. 2012; 79(2):124-130 | DOI: 10.55095/achot2012/018

Náhrada prvního metatarzofalangeálního skloubení MedinPůvodní práce

R. HROMÁDKA1,2,*, V. BARTÁK1, A. SOSNA1, S. POPELKA1
1 Ortopedická klinika 1. LF UK Praha a FN Motol
2 Anatomický ústav 1. LF UK Praha

PURPOSE OF THE STUDY:
Hemiarthoplasty or total replacement of the first metatarsophalangeal (MTP) joint has been used in orthopaedic surgery for the last 60 year, but good post-operative outcomes have been achieved only in the last ten years. Joint replacement is mainly used in stage 3 and stage 4 hallux rigidus conditions for which arthrodesis is not indicated. The operation on the first MTP joint has its place in the present-day orthopaedics. This study describes anatomical measurements and the development of the first Czech implant (MEDIN Orthopaedics) to replace this joint.

MATERIAL AND METHODS:
Thirty cadaver specimens were used to develop basic shapes of phalangeal and metatarsal components. A standard technique was used for anatomical dissection of the first MTP joint. Fifteen specimens were cut in the sagittal plane and fifteen in the transverse plane in order to open the intramedullary cavity of the proximal phalanx of the great toe and the first metatarsal bone. The basic shapes of phalangeal and metatarseal components were designed based on the shape of cortical bone of their inner surfaces. Data for the shape, size and scale of articular components were obtained by measurement on 58 dry bone specimens of the first metatarsus and on 30 calibrated X-ray images. In order to adjust the scale and size of components, the final shape and the range of implant size were tested on 50 specimens of dissected lower extremities fixed in formaldehyde, acetone, ethyl-alcohol and glycerol.

RESULTS:
The new type of a first MTP implant designed by us was based on cone-shaped anchor components coated with hydroxyapatite. The implants can be used in hemiarthroplasty or total joint replacement. The metatarsal insert was designed with a declination angle of 20 degrees to facilitate good dorsiflexion and with a flattening to ensure good function of the sesamoid bones, The phalangeal articular insert was made of either CoCr alloy or low-weight polyethylene for use in hemiarthroplasty and total joint replacement, respectively.

DISCUSSION:
The new implants are designed for treatment of stage 3 or stage 4 hallux rigidus. We recommend to use hemiarthroplasty or total joint replacement only in the case of first metatarsal head destruction or severe joint destruction due to rheumatoid arthritis.
CONSLUSIONS
Our anatomical study of the first MTP joint, proximal phalanx of the great toe and first metatarsal bone was used to design the first Czech implant of this joint.

Klíčová slova: hallux rigidus, hallux limitus, total joint replacement, hemiarthroplasty, metatarsophalangeal joint

Zveřejněno: 1. duben 2012  Zobrazit citaci

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
HROMÁDKA R, BARTÁK V, SOSNA A, POPELKA S. Náhrada prvního metatarzofalangeálního skloubení Medin. Acta Chir Orthop Traumatol Cech. 2012;79(2):124-130. doi: 10.55095/achot2012/018. PubMed PMID: 22538102.
Stáhnout citaci

Reference

  1. BARTÁK, V., HROMÁDKA, R., FULÍN, P., JAHODA, D., SOSNA, A., POPELKA, S.: Anatomická studie úponové části m. flexor hallucis brevis pro klinickou praxi. Acta Chir. orthop. Traum. čech., 78: 145-148, 2011. Přejít k původnímu zdroji...
  2. BARTÁK, V., POPELKA, S., HROMÁDKA, R., PECH, J., JAHODA, D., SOSNA, A.: Naše zkušenosti s náhradou I. MTP kloubu palce implantátem ToeFit Plus. Acta Chir. orthop. Traum. čech., 77: 222-227, 2010. Přejít k původnímu zdroji...
  3. BURNS, A.: Implant procedures. In: Gerbert, J. (eds): Textbook of bunion surgery. 2nd Ed. Futura Publishing, Mount Kisco, New-York 1991, 269.
  4. COUGHLIN, M. J., ABDO, R. V.: Arthrodesis of the first metatarsophalangeal joint with Vitallium plate fixation. Foot Ankle Int, 15: 18-28, 1994. Přejít k původnímu zdroji... Přejít na PubMed...
  5. COUGHLIN, M. J., SHURNAS, P. S.: Hallux rigidus. Grading and long term results of operative treatment. J. Bone Jt Surg, 83-A: 2072-2088, 2003. Přejít k původnímu zdroji...
  6. D'ANTONIO, J. A., CAPELLO, W. N., MANLEY, M. T.: Remodeling of bone around hydroxyapatite- coated femoral stems. J. Bone Jt Surg., 78-A: 1226-1234, 1996. Přejít k původnímu zdroji... Přejít na PubMed...
  7. ENDLER, F.: Zur entwiclung einer kuenstlichen artrhroplastik des grosszehnengrundgelenke und ihre bisheride indication. Z. Orthop., 80: 480, 1951.
  8. FILIP, L., STEHLÍK, J., MUSIL, D., SADOVSKÝ, P.: Indikace a metody léčby hallux rigidus na našem pracovišti. Acta Chir. orthop. Traum. čech., 75: 173-179, 2008. Přejít k původnímu zdroji...
  9. FRISCH, E.: Biomaterials in foot surgery. Clin. Podiatry, 1: 11, 1984. Přejít k původnímu zdroji...
  10. FUHRMANN, R. A.: MTP Prosthesis (ReFlexion) for hallux rigidus. Tech. Foot Ankle Surg., 4: 2-9, 2005. Přejít k původnímu zdroji...
  11. GEESINK, R., G., T., MANLEY, M., T.: Hydroxyapatite coating in orthopaedical surgery. New York, Raven 1993.
  12. GHALAMBOR, N., CHO, D., R., GOLDRING, S., R., NIHAL, A., TREPMAN, E.: Microscopic metallic wear and tissue response in failed titanium Hallux metatarsophalangeal implants: two cases. Foot Ankle Int., 23: 158 -162, 2002. Přejít k původnímu zdroji... Přejít na PubMed...
  13. GIANNINI, S., MORONI, A.: Alumina total joint replacement of the first metatarsophalangeal joint. In: Bonfield, W., Hastings, G., W., Tanner, K., E. (eds): Bioceramics.Vol. 4. Butterworth-Heinemann, London 1991, 39. Přejít k původnímu zdroji...
  14. JARVIS, B. D., MOATS, D. B., BURNS, A., GERBERT, J.: Lawrence design first metatarsophalangeal joint prosthesis. J. Am. Podiatr. Med. Assoc., 76: 617, 1986. Přejít k původnímu zdroji... Přejít na PubMed...
  15. JENKIN, W. M., OLOF, L. M.: Implant arthroplasty in the rheumatoid arthritic patient. Clin. Podiatr. Med. Surg., 5: 213, 1988. Přejít k původnímu zdroji...
  16. JOHNSON, K. A., BUCK, P. G.: Total replacement arthroplasty of the first metatarsophalangeal joint. Foot Ankle, 1: 307, 1981. Přejít k původnímu zdroji... Přejít na PubMed...
  17. JOPLIN, R. J.: The proper digital nerve, vitallium stem arthroplasty, and some thoughts about foot surgery in general. Clin. Orthop. Relat. Res., 76: 199, 1971. Přejít k původnímu zdroji... Přejít na PubMed...
  18. KELLER, W. L.: The surgical treatment of bunion and hallux valgus. N. Y. Med. J., 80: 741, 1904.
  19. KOFOED, H.: Alternatives to fusion. The evidence. In: EFAS advance forefoot symposium. 11.-12. December 2009.
  20. LANDOR, I., VAVŘÍK, P., JAHODA, D., POKORNÝ, D., BALLAY, R., SOSNA, A.: Dlouhodobé zkušenosti s kombinovaným hydroxyapatitovým povrchem ARBOND v osteointegraci implantátu. Acta Chir. ortop. Traum. čech., 76: 172-178, 2009. Přejít k původnímu zdroji...
  21. LANGE, J., MERK, H., BARZ, T., WALTHER, C., FOLLAK, N.: Mittelfristige Ergebnise der ToeFit-Grosszehengrundgelenksendoprothese. Z. Orthop. Unfall, 146: 609-615, 2008. Přejít k původnímu zdroji... Přejít na PubMed...
  22. LUCY, R.: Stiffness of the great toe in male adolescents. Brit. Med. J 1: 726-727, 1887. Přejít k původnímu zdroji...
  23. MOUZIN, O., SOBALLE, K., BECHTOLD, J. E.: Loading improves anchorage of hydroxyapatite implants more than titanium implants. J. Biomed. Mat. Res. Appl. Biomater., 58: 61-68, 2001. Přejít k původnímu zdroji...
  24. POKORNÝ, D., ŠLOUF, M., VESELÝ, F., FULÍN, P., JAHODA, D., SOSNA, A.: Distribuce otěrových částic UHMWPE v periprotetických tkáních u TEP kyčelního kloubu. Acta Chir. orthop. Traum. čech., 77: 87-92, 2010. Přejít k původnímu zdroji...
  25. PULAVARTI, R. S., McVIE, J. L., TULLOCH, C. J.: First Metatarsophalangeal joint replacement using the bio-Action great toe implant: Intermediate Results. Foot Ankle Int., 26: 1033-1038, 2005. Přejít k původnímu zdroji... Přejít na PubMed...
  26. RZONCA, E., LEVITZ, S., LUE, B.: Hallux equinus. J. Am. Pediatry Assoc., 74: 390- 393, 1984. Přejít k původnímu zdroji... Přejít na PubMed...
  27. SEEBER, E., KNESSL, J.: Die Grosszehengrundgelenksprothese "Toefit-Plus". Med. Orthop. Tech., 122: 98-100, 2002.
  28. SEEBER, E., KNESSL, J.: Treatment of hallux rigidus with the ToeFit-Plus joint replacement system. Interact. Surg., 2: 77-85, 2007. Přejít k původnímu zdroji...
  29. SEEBURGER, R. H.: Surgical implants of alloyed metal in joints of the feet. J. Am. Podiatry Assoc., 54: 391, 1964.
  30. SORBIE, CH., SAUNDERS, G. A. B.: Hemiartroplasty in the treatment of hallux rigidus. Foot Ankle Int., 29: 273-281, 2008. Přejít k původnímu zdroji... Přejít na PubMed...
  31. SUN, L., BERNDT, C. C., GROSS, K. A., KUCUK, A.: Material fundamentals and clinical performance of plasma-sprayed hydroxyapatite coatings: a review. J. Biomed. Mat. Res. Appl. Biomater., 58: 570-592, 2001. Přejít k původnímu zdroji... Přejít na PubMed...
  32. SWANSON, A. B.: Flexible implant resektion arthroplasty in the hand and extremities. St. Luis, CV Mosby 1973.
  33. ŠVARC, A., PILNÝ, J., KUBEŠ, J.: Naše zkušenosti s operační rekonstrukcí přednoží u pacientů s hallux valgus a metatarzalgiemi při pes transversoplanus. Acta Chir. orthop. Traum. čech., 77(5): 432-435, 2010. Přejít k původnímu zdroji...
  34. WEIL, L. S., POLLAK, R. A., GOLLER, W. L.: Total first joint replacement in hallux valgus and hallux rigidus, longterm results in 484 cases. Clin. Podiatry, 1: 103, 1984. Přejít k původnímu zdroji...