Acta Chir Orthop Traumatol Cech. 2012; 79(4):317-323 | DOI: 10.55095/achot2012/045
TEP kyčle Plasmacup-Bicontact: výsledky s minimálně 10letým sledovánímPůvodní práce
- 1 Ortopedická klinika LF UP a FN Olomouc
- 2 Laboratoř experimentální medicíny, Ústav molekulární a translační medicíny, LF UP a FN Olomouc
PURPOSE OF THE STUDY:
The aim of the study was to analyse the long-term clinical and radiographic results of total hip arthroplasty (THA) with the Bicontact-Plasmacup prosthesis.
MATERIAL AND METHODS:
In this retrospective observational study, 34 consecutive patients undergoing Bicontact-Plasmacup THA between August 1998 and July 2000 were evaluated. The group included 15 women and 19 men with an average age of 56 years (30 to 67 years). The mean ± SD of follow-up was 132 ± 9 months (122 to 149 months). Patients' satisfaction with the treatment outcome and selected clinical and radiographic features were evaluated. The data were analysed using the concordance test, Spearman's correlation coefficient and chi-square test.
RESULTS:
At the final follow-up, 94% of the interviewed patients reported satisfaction with the surgery outcome. The mean Harris score was 93 (67 to 98) points, with excellent or very good results in 91% of the patients. Compared with the post-operative radiographs, the mean ± SD value for vertical migration of the acetabular cup was 2.43 ± 3.21 mm (0 to 12 mm) and that for horizontal migration was 0.35 ± 1.0 mm (0 to 4.3 mm). The mean stem subsidence was by 3.87 mm (range, 0.36 to 21.11 mm; SD 4.03). At the final follow-up, all implants showed radiographic stability and absence of radiolucent lines. Early acetabular osteolysis was detected in two patients (5.9 %). In 19 patients (56%) a change in bone architecture, similar to alterations due to weight-bearing adaptation, was seen around the cup following surgery. The stress shielding effect of the femoral component was recorded in 79%, bypass phenomenon in 41 % and pedestal formation under the apex of the stem in 18% of the patients. The mean rate of polyethylene wear measured by the Dorr and Wan method was 0.106 mm per year (range, 0.00 to 0.267 mm) or it was 0.078 mm per year (0.00 to 0.19 mm) when assessed by a modification of the method described by Kang et al.
DISCUSSION:
The clinical and radiographic findings presented here are in accordance with the relevant literature data. This implant design is associated with a relatively infrequent osteolysis around the cup probably due to a low wear rate of polyethylene and a stable polyethylene liner-metal cup interface. Based on this work and some other studies it can be concluded that a post-operative migration of several millimetres is compatible with a long-term, stable cementless THA.
CONCLUSIONS:
The Bicontact-Plasmacup THA meets the National Institute for Health and Clinical Excellence (NICE) criteria for a very good implant (ten-year survival rate of over 90%). At 10 years after surgery, however, the first signs of complaints about surgery outcome // dissatisfaction with surgery outcomes were recorded. This may, at least partly, be related to aseptic loosening which is one of the symptoms of particle disease.
Klíčová slova: total hip arthroplasty, Plasmacup-Bicontact prosthesis
Zveřejněno: 1. srpen 2012 Zobrazit citaci
Reference
- ASPENBERG, P., WAGNER, P., NILSSON, K. G., RANSTAM, J.: Fixed or loose? Dichotomy in RSA data for cemented cups. Acta Orthop., 79: 467-473, 2008.
Přejít k původnímu zdroji...
Přejít na PubMed...
- BELLAMY, N., WILSON, C., HENDRIKZ, J., WHITEHOUSE, S. L., PATEL, B., DENNISON, S., DAVIS, T.: Osteoarthritis Index delivered by mobile phone (m-WOMAC) is valid, reliable, and responsive. J. Clin. Epidemiol., 64: 182-190, 2011.
Přejít k původnímu zdroji...
Přejít na PubMed...
- BROOKER, A. F., BOWERMAN, J. W., ROBINSON, R. A., RILEY, L. H., Jr.: Ectopic ossification following total hip replacement. Incidence and a method of classification. J. Bone Jt Surg., 55-A: 1629-1632, 1973.
Přejít k původnímu zdroji...
- DELEE, J. G., CHARNLEY, J.: Radiological demarcation of cemented sockets in total hip replacement. Clin. Orthop. Relat. Res., 121: 20-32, 1976.
Přejít k původnímu zdroji...
- EL-WARRAK, A. O., OLMSTEAD, M., SCHNEIDER, R., MEINEL, L., BETTSCHART-WOLFISBERGER, R., AKENS, M. K., AUER, J., VON RECHENBERG, B.: An experimental animal model of aseptic loosening of hip prostheses in sheep to study early biochemical changes at the interface membrane. BMC Musculoskelet. Disord., 5: 7, 2004.
Přejít k původnímu zdroji...
Přejít na PubMed...
- EMMS, N. W., STOCKLEY, I., HAMER, A. J., WILKINSON, J. M.: Long-term outcome of a cementless, hemispherical, press-fit acetabular component: survivorship analysis and dose-response relationship to linear polyethylene wear. J. Bone Jt Surg., 92-B: 856-861, 2010.
Přejít k původnímu zdroji...
Přejít na PubMed...
- ENGH, C. A., MASSIN, P., SUTHERS, K. E.: Roentgenographic assessment of the biologic fixation of porous-surfaced femoral components. Clin. Orthop. Relat. Res., 257: 107-128, 1990.
Přejít k původnímu zdroji...
- GALLO, J., HAVRANEK, V., ZAPLETALOVA, J.: Risk factors for accelerated polyethylene wear and osteolysis in ABG I total hip arthroplasty. Int. Orthop., 34: 19-26, 2010.
Přejít k původnímu zdroji...
Přejít na PubMed...
- GALLO, J., KAMÍNEK, P., ZAPLETALOVÁ, J.; ČECHOVÁ, I., ©PIČKA, J., DITMAR, R.: Je osteolýza kolem stabilní necementované TEP kyčle asymptomatická? Acta Chir. orthop. Traum. čech., 71: 20-25, 2004.
- GOSSEC, L., HAWKER, G., DAVIS, A. M., MAILLEFERT, J. F., LOHMANDER, L. S., ALTMAN, R., CIBERE, J., CONAGHAN, P. G. et al.: OMERACT/OARSI initiative to define states of severity and indication for joint replacement in hip and knee osteoarthritis. J. Rheumatol., 34: 1432-1435, 2007.
- GRUEN, T. A., McNEICE, G. M., AMSTUTZ, H. C.: "Modes of failure" of cemented stem-type femoral components: a radiographic analysis of loosening. Clin. Orthop. Relat. Res., 141: 17-27, 1979.
Přejít k původnímu zdroji...
- HAWKER, G. A., GIGNAC, M. A., BADLEY, E., DAVIS, A. M., FRENCH, M. R., LI, Y., PERRUCCIO, A. V., POWER, J. D., SALE, J., LOU, W.: A longitudinal study to explain the paindepression link in older adults with osteoarthritis. Arthritis Care Res., 63: 1382-1390, 2011.
Přejít k původnímu zdroji...
Přejít na PubMed...
- HOOGEBOOM, T. J., DEN BROEDER, A. A., SWIERSTRA, B. A., DE BIE, R. A., VAN DEN ENDE, C. H.: Joint-pain comorbidity is associated with unfavorable health status and medication use in hip and knee osteoarthritis: A cross-sectional study. Arthritis Care Res., [Epub ahead of print 2011 Sep 27], 64: 54-58,2012.
Přejít k původnímu zdroji...
Přejít na PubMed...
- KANG, J. S., PARK, S. R., EBRAMZADEH, E., DORR, L. D.: Measurement of polyethylene wear in total hip arthroplasty-accuracy versus ease of use. Yonsei Med. J., 44: 473-478, 2003.
Přejít k původnímu zdroji...
Přejít na PubMed...
- KROELL, A., BEAULE, P., KRISMER, M., BEHENSKY, H., STOECKL, B., BIEDERMANN, R.: Aseptic stem loosening in primary THA: migration analysis of cemented and cementless fixation. Int. Orthop., 33: 1501-1505, 2009.
Přejít k původnímu zdroji...
Přejít na PubMed...
- LANGOVA, K., GALLO, J.: Je Kaplan-Meierova statistika nejvhodnějąí metodou k hodnocení přeľívání výsledku v ortopedii? Acta Chir. orthop. Traum. čech., 77: 118-123, 2010.
Přejít k původnímu zdroji...
- MJOBERG, B.: Fixation and loosening of hip prostheses. A review. Acta Orthop. Scand., 62: 500-508, 1991.
Přejít k původnímu zdroji...
Přejít na PubMed...
- PAGNANO, W., HANSSEN, A. D., LEWALLEN, D. G., SHAUGHNESSY, W. J.: The effect of superior placement of the acetabular component on the rate of loosening after total hip arthroplasty. J. Bone Jt Surg., 78-A: 1004-1014, 1996.
Přejít k původnímu zdroji...
Přejít na PubMed...
- PARSONS, N. R., DE SOUZA, R. M., ONI, T., ACHTEN, J., KRIKLER, S. J., COSTA, M. L.: A comparison of Harris and Oxford hip scores for assessing outcome after resurfacing arthroplasty of the hip: can the patient tell us everything we need to know. Hip Int., 20: 453-459, 2010.
Přejít k původnímu zdroji...
Přejít na PubMed...
- POSPISCHILL, M., KNAHR, K.: Cementless total hip arthroplasty using a threaded cup and a rectangular tapered stem. Follow-up for ten to 17 years. J. Bone Jt Surg., 87-B: 1210-1215, 2005.
Přejít k původnímu zdroji...
Přejít na PubMed...
- RITTER, M. A., THONG, A. E., DAVIS, K. E., BEREND, M. E., MEDING, J. B., FARIS, P. M.: Long-term deterioration of joint evaluation scores. J. Bone Jt Surg., 86-B: 438-442, 2004.
Přejít k původnímu zdroji...
Přejít na PubMed...
- RODER, C., STAUB, L. P., EICHLER, P., WIDMER, M., DIETRICH, D., EGGLI, S., MULLER, U.: Avoiding misclassification bias with the traditional Charnley classification: rationale for a fourth Charnley class BB. J. Orthop. Res., 24: 1803-1808, 2006.
Přejít k původnímu zdroji...
Přejít na PubMed...
- SCHMALZRIED, T. P., SHEPHERD, E. F., DOREY, F. J., JACKSON, W. O., DELA ROSA, M., FA'VAE, F., McKELLOP, H. A., McCLUNG, C. D., et al.: The John Charnley Award. Wear is a function of use, not time. Clin. Orthop. Relat. Res., 381: 36-46, 2000.
Přejít k původnímu zdroji...
Přejít na PubMed...
- SWAMY, G., PACE, A., QUAH, C., HOWARD, P.: The Bicontact cementless primary total hip arthroplasty: long-term results. Int. Orthop., [Epub ahead of print 2010 Sept 10], 36: 915-920, 2012.
Přejít k původnímu zdroji...
Přejít na PubMed...
- TERWEE, C. B., BOUWMEESTER, W., VAN ELSLAND, S. L., DE VET, H. C., DEKKER, J.: Instruments to assess physical activity in patients with osteoarthritis of the hip or knee: a systematic review of measurement properties. Osteoarthritis Cartilage, 19: 620-633, 2011.
Přejít k původnímu zdroji...
Přejít na PubMed...
- UTTING, M. R., LANKESTER, B. J., SMITH, L. K., SPENCER, R. F.: Total hip replacement and NICE. BMJ, 330: 318-319, 2005.
Přejít k původnímu zdroji...
Přejít na PubMed...
- WAMPER, K. E., SIEREVELT, I. N., POOLMAN, R. W., BHANDARI, M., HAVERKAMP, D.: The Harris hip score: Do ceiling effects limit its usefulness in orthopedics? Acta Orthop., 81: 703-707, 2010.
Přejít k původnímu zdroji...
Přejít na PubMed...
- WAN, Z., DORR, L. D.: Natural history of femoral focal osteolysis with proximal ingrowth smooth stem implant. J. Arthroplasty, 11: 718-725, 1996.
Přejít k původnímu zdroji...
Přejít na PubMed...
- ZICAT, B., ENGH, C. A., GOKCEN, E.: Patterns of osteolysis around total hip components inserted with and without cement. J. Bone Jt Surg., 77-A: 432-439, 1995.
Přejít k původnímu zdroji...
Přejít na PubMed...