PURPOSE OF THE STUDY:
The aim of our study is to solve the problem of insufficient fixation of comminuted fractures of the quadrilateral plane and the iliopectineal line. These fixation problems occur while using the standard narrow 3.5 mm fixation plate applied from a modified Stoppa approach. A new plate developed by the authors - the Omega plate - fulfils the requirements.

MATERIAL AND METHODS:
In the period 2010-2012, we performed 156 stabilisations of pelvic ring fractures and acetabular fractures. We used the modified Stoppa approach applying the standard fixation plate in 24 patients and the Omega plate in 15 patients. The patient group with the Omega plate included 10 male and five female patients with the average age of 61 years (range, 30-72). Only 11 patients were followed up, with an average period of 13.3 months, because one patient was lost to follow-up and three patients were shortly after surgery. The surgical technique of Omega plate application is described in detail. The clinical evaluation of post-operative results was based on the Harris Hip Score; the graphical results were rated using the Matta and Pohlemann criteria.

RESULTS:
The Stoppa approach alone was used in four patients, combination of two approaches (Stoppa and Kocher-Langenbeck approach) was used in six cases and three approaches were employed in five patients. No adverse intra- or post-operative events were recorded. Excellent or satisfactory graphical results were obtained in 12 patients and an unsatisfactory graphical outcome was recorded in three cases. In the follow-up period ranging from 8 to 22 months, 11 patients healed. Late complications included avascular femoral head necrosis in two and severe post-traumatic coxarthrosis in three patients. Due to these complications, all five patients underwent total hip arthroplasty without previous Omega plate removal at an average interval of 15 months from the primary pelvic surgery. They were not included in the follow-up evaluation. The remaining six patients had an average Harris Hip Score of 88 points (range, 81-98).

DISCUSSION:
The novel plate, shaped as a reverse omega letter (Ω), enables fixation of the quadrilateral area of the acetabulum through pressure of the arc of the plate against this area. Hitches, with holes for screw insertion, attached to the Omega plate in its middle part allow for fixation of fragments above the linea arcuata simply by pressure. Hitches in the ventral part provide for plate fixation to the ventral acetabular column and the superior pubic ramus. Hitches in the posterior segment of the plate facilitate insertion of a long screw in the posterior acetabular column from an additional iliac approach for stabilisation of simple acetabular fractures. The Omega plates are manufactured in several modifications.

CONCLUSIONS:
The Omega plate enables us to fix fractures of the superior pubic ramus, fractures of the anterior acetabular column, fractures of the quadrilateral acetabular plate, fractures in the iliopectineal line and simple fractures of the posterior column. A CT-defined projection of the pelvic inlet based on pre-operative CT scans allows us to choose the appropriate plate size and to shape the plate pre-operatively. After a technically well performed Stoppa approach and good fragment reduction, the application of an Omega plate is easy if our recommendations are followed. Fixation of all fragments of the anterior column and the quadrilateral plate is very stable and the Omega plate is highly resistant to secondary loss of reduction. A potential total hip arthroplasty does not require Omega plate removal.

Keywords: acetabular fracture; modified Stoppa approach; Omega plate; CT-defined pelvic inlet projection

Published: April 1, 2013  Show citation

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
Šrám J, Taller S, Lukáš R, Endrych L. Use of the Omega Plate for Stabilisation of Acetabular Fractures: First Experience. Acta Chir Orthop Traumatol Cech. 2013;80(2):118-124. doi: 10.55095/achot2013/018. PubMed PMID: 23562255.
Download citation

References

  1. ANDERSEN, R. C., O'TOOLE, R. V., NASCONE, J. W., SCIADINI, M. F., FRISCH, H. M., TUREN, C. W.: Modified Stoppa approach for acetabular fractures with anterior and posterior column displacement: quantification of radiographic reduction and analysis of interobserver variability. J. Orthop. Trauma, 24: 271-278, 2010. Go to original source... Go to PubMed...
  2. COLE, J. D., BOLHOFNER, B. R.: Acetabular fracture fixation via a modified Stoppa limited intrapelvic approach: description of operative technique and preliminary treatment results. Clin. Orthop., 305: 112-123, 1994. Go to original source...
  3. HARRIS, W. H.: Traumatic arthritis of the hip after dislocation and acetabular fractures: treatment by mold arthroplasty. An end-result study using a new method of result evaluation. J. Bone Jt Surg., 51-A: 737-755, 1969. Go to original source...
  4. HIRVENSALO, E., LINDAHL, J., BOSTMAN, O.: A new approach to the internal fixation of unstable pelvic fractures. Clin. Orthop., 297: 28-32, 1993. Go to original source...
  5. KACRA, B. K., ARAZI, M., CICEKCIBASI, A. E., BUYUKMUMCU, M., DEMIRCI, S.: Modified medial Stoppa approach for acetabular fractures: an anatomic study. J. Trauma, 71: 1340-1344, 2011. Go to original source... Go to PubMed...
  6. MAST, J., JAKOB, R., GANZ, R.: Planning and reduction technique in fracture surgery. New York, Springer-Verlag, 1989. Go to original source...
  7. MATTA, J. M.: Operative treatment of acetabular fractures through the ilioinguinal approach: a 10-year perspective. J. Orthop. Trauma, 20: S20-S29, 2006. Go to original source... Go to PubMed...
  8. McERLAIN, M., KHAN, O., WARD, A., CHESSER, T.: The use of the Stoppa approach in the operative treatment of pelvic and acetabular trauma. J. Bone Jt Surg., 88-B (Suppl. 2): 264, 2006.
  9. PAVELKA, T., HOUČEK, P.: Komplikace operačního léčení zlomenin acetabula. Acta Chir. orthop. Traum. čech., 76: 186-193, 2009. Go to original source...
  10. POHLEMANN, T., GÄNSSLEN, A., SCHELLWALD, O., CULEMANN, U., TSCHERNE, H.: Outcome after pelvic ring injuries. Injury, 27 (Suppl. 3): B31-B38, 1996. Go to original source... Go to PubMed...
  11. PONSEN, K. J., JOOSSE, P., SCHIGT, A., GOSLINGS, C. J., LUITSE, J. S.: Internal fracture fixation using the Stoppa approach in pelvic ring and acetabular fractures: technical aspects and operative results. J. Trauma, 61: 662-667, 2006. Go to original source... Go to PubMed...
  12. SAGI, H. C., AFSARI, A., DZIADOSZ, D.: The anterior intra-pelvic (modified rives-stoppa) approach for fixation of acetabular fractures. J. Orthop. Trauma, 24: 263-270, 2010. Go to original source... Go to PubMed...
  13. SHAZAR, N., MOSHEIFF, R., BERNSTEIN-WEYEL, M., BRUCK, N., KHOURY, A.: The Stoppa approach for acetabular fractures. J. Bone Jt Surg., 90-B (Suppl. 3): 517, 2008.
  14. STOPPA, R. E., RIVES, J. L., WARLAUMONT, C. R., PALOT, J. P., VERHAEGHE, P. J., DELATTRE, J. F.: The use of Dacron in the repair of hernias of the groin. Surg. Clin. North Am., 64: 269-285, 1984. Go to original source... Go to PubMed...
  15. ŠRÁM, J.: New plate for reconstruction quadrilateral surface of the acetabulum. In: ECTES 2012. 13th European Congress of Trauma & Emergeny Surgery. Many ways-one goal. Basel, May 12-15, 2012.
  16. TALLER, S., ŠRÁM, J., LUKÁŠ, R., KŘIVOHLÁVEK, M.: Zlomeniny pánevního kruhu a acetabula operované přístupem dle Stoppy. Acta Chir. orthop. Traum. čech., 77: 93-98, 2010. Go to original source...
  17. TILE, M., HELFET, D. L., KELLAM, J. E.: Fractures of the pelvis and acetabulum. Third edition. Baltimore, Lippincott Williams & Wilkins, 2003.