Acta Chir Orthop Traumatol Cech. 2013; 80(6):377-385 | DOI: 10.55095/achot2013/064
What Should an Orthopaedic Surgeon Know about Biotribology of Total Hip ArthroplastyCurrent concepts review
- 1 Ortopedická klinika LF UP a FN Olomouc
- 2 Ústav konstruování, Fakulta strojního inženýrství, VUT v Brně
The objective of this paper is to briefly review the engineering biotribology of total hip arthroplasty (THA). At present the issue of lubrication, wear and friction within a THA is being studied in detail and the results have been published in a number of papers. The reason is obvious since the rate at which the articulating surfaces wear off affects the longevity of a THA to a great extent. Seeking an optimal biotribological configuration remains an important task for both the researchers and orthopaedic surgeons. In order to enable the orthopaedic community to think over THA biotribology and, at the same time, communicate with the representatives of technical professions, knowledge of the basic principles of biotribology and their understanding is necessary.
Keywords: total hip arthroplasty (THA), tribology, biotribology, bearing materials, friction, lubrication, wear, osteolysis, aseptic loosening
Published: December 1, 2013 Show citation
ACS | AIP | APA | ASA | Harvard | Chicago | Chicago Notes | IEEE | ISO690 | MLA | NLM | Turabian | Vancouver |
References
- ADELANI, M. A., KEENEY, J. A., PALISCH, A., FOWLER, S. A., CLOHISY, J. C.: Has total hip arthroplasty in patients 30 years or younger improved? A systematic review. Clin. Orthop. Relat. Res., 471: 2595-2601, 2013.
Go to original source...
Go to PubMed...
- AFFATATO, S., TRAINA, F., TONI, A.: Microseparation and stripe wear in alumina-on-alumina hip implants. Int. J. Artif. Org., 34: 506-512, 2011.
Go to original source...
Go to PubMed...
- ANDERSSON, M. K., ANISSIAN, L., STARK, A., BUCHT, E., FELLANDER-TSAI, L., TSAI, J. A.: Synovial fluid from loose hip arthroplasties inhibits human osteoblasts. Clin. Orthop. Relat. Res., 378: 148-154, 2000.
Go to original source...
Go to PubMed...
- BASHINSKAYA, B., ZIMMERMAN, R. M., WALCOTT, B. P., ANTOCI, V.: Arthroplasty utilization in the United States is predicted by age-specific population groups. ISRN orthopedics. Open access: 2012.
Go to original source...
- BERGMANN, G., GRAICHEN, F., ROHLMANN, A., BENDER, A., HEINLEIN, B., DUDA, G. N., HELLER, M. O., MORLOCK, M. M.: Realistic loads for testing hip implants. Bio-med. Mater. Engin.,20: 65-75, 2010.
Go to original source...
Go to PubMed...
- BERNTHAL, N. M., CELESTRE, P. C., STAVRAKIS, A. I., LUDINGTON, J. C., OAKES, D. A.: Disappointing short-term results with the DePuy ASR XL metal-on-metal total hip arthroplasty. J. Arthroplasty, 27: 539-544, 2012.
Go to original source...
Go to PubMed...
- BEZNOSKA, S., ČECH, O., LÖBL, K.: Umělé náhrady lidských kloubů. Praha, SNTL; 1987.
- BRAND, R. A.: Joint contact stress: a reasonable surrogate for biological processes? Iowa Orthop. J., 25: 82-94, 2005.
- BROCKETT, C., WILLIAMS, S., JIN, Z., ISAAC, G., FISHER, J.: Friction of total hip replacements with different bearings and loading conditions. J. Biomed. Mater. Res. Part B, App. Biomat., 81: 508-515, 2007.
Go to original source...
Go to PubMed...
- BROCKETT, C. L., WILLIAMS, S., JIN, Z., ISAAC, G. H., FISHER, J.: Squeaking hip arthroplasties: a tribological phenomenon. J. Arthroplasty, 28: 90-97, 2013.
Go to original source...
Go to PubMed...
- COOKE, A. F., DOWSON, D., WRIGHT, V.: The pressure-viscosity characteristics of synovial fluid. Biorheology, 15: 129-135, 1978.
Go to original source...
Go to PubMed...
- DOWSON, D.: New joints for the Millennium: wear control in total replacement hip joints. Proc. Instit. Mech. Eng. Part H., 215: 335-358, 2001.
Go to original source...
Go to PubMed...
- DOWSON, D.: Bio-tribology. Faraday Discuss., 156: 9-30; 2012.
Go to original source...
Go to PubMed...
- FISHER, J.: Bioengineering reasons for the failure of metal-on-metal hip prostheses: an engineer's perspective. J. Bone Jt. Surg., 93-B: 1001-1004, 2011.
Go to original source...
Go to PubMed...
- FISHER, J.: A stratified approach to pre-clinical tribological evaluation of joint replacements representing a wider range of clinical conditions advancing beyond the current standard. Faraday Discuss., 156: 59-68, 2012.
Go to original source...
Go to PubMed...
- FISHER, J., JIN, Z., TIPPER, J., STONE, M., INGHAM, E.: Tribology of alternative bearings. Clin. Orthop. Relat. Res., 453: 25-34, 2006.
Go to original source...
Go to PubMed...
- GALLO, J., GOODMAN, S. B., KONTTINEN, Y. T., RASKA, M.: Particle disease: biologic mechanisms of periprosthetic osteolysis in total hip arthroplasty. Innate immunity, 19: 213-224, 2013.
Go to original source...
Go to PubMed...
- GALLO, J., GOODMAN, S. B., LOSTAK, J., JANOUT, M.: Advantages and disadvantages of ceramic on ceramic total hip arthroplasty: a review. Biomed. Papers Medic. Fac. University Palack., 156: 204-212, 2012.
Go to original source...
Go to PubMed...
- GALLO, J., LANGOVA, K., HAVRANEK, V., CECHOVA, I.: Poor survival of ABG I hip prosthesis in younger patients. Biomed. Papers Medic. Fac. University Palack.,152: 163-168, 2008.
- GISPERT, M. P., SERRO, A. P., COLACO, R., SARAMAGO, B.: Friction and wear mechanisms in hip prosthesis: Comparison of joint materials behaviour in several lubricants. Wear, 260: 149-158, 2006.
Go to original source...
- GOLDSMITH, A. A., DOWSON, D., WROBLEWSKI, B. M., SINEY, P. D., FLEMING, P. A., LANE, J. M., STONE, M. H., WALKER, R.: Comparative study of the activity of total hip arthroplasty patients and normal subjects. J. Arthroplasty, 16: 613-619, 2001.
Go to original source...
Go to PubMed...
- HAMROCK, B. J., DOWSON, D.: Elastohydrodynamic lubrication of elliptical contacts for materials of low elastic-modulus I - Fully flooded conjunction. J. Lubric. Tech.-T. ASME, 100: 236-245, 1978.
Go to original source...
- HORÁK, Z., POKORNÝ, D,, FULÍN, P., ŠLOUF, M., JAHODA, D., SOSNA, A.: Polyetheretherketone (PEEK). Část 1: perspektivní materiál pro ortopedickou a traumatologickou praxi. Acta Chir. orthop. Traum. čech., 77: 463-469, 2010.
Go to original source...
- ISAAC, G. H., THOMPSON, J., WILLIAMS, S., FISHER, J.: Metal-on-metal bearings surfaces: materials, manufacture, design, optimization, and alternatives. Proc. Instit. Mech. Eng. Part H, 220: 119-133, 2006.
Go to original source...
Go to PubMed...
- JASTY, M., RUBASH, H. E., MURATOGLU, O.: Highly cross-linked polyethylene: the debate is over-in the affirmative. J. Arthroplasty,20: 55-58, 2005.
Go to original source...
Go to PubMed...
- JIN, Z. M., STONE, M., INGHAM, E., FISHER, J.: Biotribology. Curr. Orthop., 20: 32-40, 2006.
Go to original source...
- JOST, P.: Tribology: How a word was coined 40 years ago. Tribol. Lubr. Technol., 62: 24-28, 2006.
- LACHIEWICZ, P. F., HECKMAN, D. S., SOILEAU, E. S., MANGLA, J., MARTELL, J. M.: Femoral head size and wear of highly cross-linked polyethylene at 5 to 8 years. Clin. Orthop. Relat. Res., 467: 3290-3296, 2009.
Go to original source...
Go to PubMed...
- LAMONTAGNE, M., BEAULIEU, M. L., VARIN, D., BEAULE, P. E.: Gait and motion analysis of the lower extremity after total hip arthroplasty: what the orthopedic surgeon should know. Orthop. Clin. North Am., 40: 397-405, 2009.
Go to original source...
Go to PubMed...
- MATTEI, L., DI PUCCIO, F., PICCIGALLO, B., CIULLI, E.: Lubrication and wear modelling of artificial hip joints: A review. Tribol. Int., 44:532-549, 2011.
Go to original source...
- MAVRAKI, A., CANN, P. M.: Friction and lubricant film thickness measurements on simulated synovial fluids. Proc. Instit. Mech. Eng. Part J - J. Eng. Tribol., 223: 325-335, 2009.
Go to original source...
- MURATOGLU, O. K., BRAGDON, C. R., O'CONNOR, D., PERINCHIEF, R. S., ESTOK, D. M., JASTY, M., HARRIS, W. H.: Larger diameter femoral heads used in conjunction with a highly cross-linked ultra-high molecular weight polyethylene: a new concept. J. Arthroplasty, 16: 24-30, 2001.
Go to original source...
Go to PubMed...
- NASSUTT, R., WIMMER, M. A., SCHNEIDER, E., MORLOCK, M. M.: The influence of resting periods on friction in the artificial hip. Clin. Orthop. Relat. Res., 407: 127-138, 2003.
Go to original source...
Go to PubMed...
- PETRTÝL, M., LÍSAL, J., DANEŠOVÁ, J., WOHLMUTHOVÁ, M.: Reologické vlastnosti synoviální tekutiny. Pohybové ústrojí: Pokroky ve výzkumu, diagnostice a terapii, 18: 210-228, 2011.
- POKORNÝ, D., ŠLOUF, M., FULÍN, P.: Současné poznatky o vlivu technologie výroby a sterilizace na strukturu, vlastnosti a životnost UHMWPE v kloubních náhradách. Acta Chir. orthop. Traum. čech., 79: 213-221, 2012.
Go to original source...
- PRENTICE, J. R., CLARK, M. J., HOGGARD, N., MORTON, A. C., TOOTH, C., PALEY, M. N., STOCKLEY, I., HADJIVASSILIOU, M., WILKINSON, J. M.: Metal-on-metal hip prostheses and systemic health: A cross-sectional association study 8 years after implantation. PloS one, 8: e66186, 2013.
Go to original source...
Go to PubMed...
- SILVA, M., MCCLUNG, C. D., DELA ROSA, M. A., DOREY, F. J., SCHMALZRIED, T. P.: Activity sampling in the assessment of patients with total joint arthroplasty. J. Arthroplasty, 20: 487-491, 2005.
Go to original source...
Go to PubMed...
- SMITH, A. J., DIEPPE, P., VERNON, K., PORTER, M., BLOM, A. W.: Failure rates of stemmed metal-on-metal hip replacements: analysis of data from the National Joint Registry of England and Wales. Lancet, 379: 1199-1204, 2012.
Go to original source...
Go to PubMed...
- VRBKA, M., Návrat, T., KŘUPKA, I., HARTL, M., ŠPERKA, P., GALLO, J.: Study of film formation in bovine serum lubricated contacts under rolling/sliding conditions. Proc. Instit. Mech. Eng. Part J, 459-475, 2013.
Go to original source...
- YAO, J. Q., LAURENT, M. P., JOHNSON, T. S., BLANCHARD, C. R., CROWNINSHIELD, R. D.: The influences of lubricant and material on polymer/CoCr sliding friction. Wear, 255: 780-784, 2003.
Go to original source...