Acta Chir Orthop Traumatol Cech. 2022; 89(1):75-80 | DOI: 10.55095/achot2022/010

Spolehlivost hodnocení na operatérovi závislých faktorů, jež ovlivňují mechanické selhání v léčbě intertrochanterické zlomeniny femuruPůvodní práce

A. TURGUT*, M. FILIBELI, M. KUMBARACI, E. BILGIN, S. HANCIOĞLU, Ö. KALENDERER
Tepecik Training and Research Hospital, Department of Orthopaedics and Traumatology, İzmir, Turkey

PURPOSE OF THE STUDY:
The intertrochanteric femur fractures seen in the elderly population are usually treated surgically. Mechanical failure of this treatment is a serious complication and to prevent this, many assessment factors have been described such as tip-apex distance, reduction quality etc. The aim of this study is to evaluate the intra- and inter-observer reliability of modifiable factors evaluated after the treatment of intertrochanteric fractures.

MATERIAL AND METHODS:
The early post-operative radiographs of sixty-two patients treated with PFN-A were evaluated. Six observers with 8 to 27 years of experience in hip trauma surgery were asked to measure tip apex distance (TAD), calcar referenced TAD (calTAD), collo-diaphyseal angle (CDA) and evaluate the position of helical blade and also fracture reductions according to Baumgaertner and Chang criteria on these radiographs. Fleiss kappa and intra-class correlation coefficient (ICC) values were calculated for inter-observer and intra-observer reliability assessment, respectively.

RESULTS:
Inter-observer reliability values for both evaluations were moderate (fleiss κ: 0.417-0.455) for TAD measurements, moderate (fleiss κ: 0.418-0.458) for calTAD measurements, fair (fleiss κ: 0.302-0.288) for CDA measurements, substantial (fleiss κ: 0.606-0.631) for antero-posterior zone evaluation, moderate (fleiss κ: 0.550-0.546) for lateral zone assessment, fair-moderate (fleiss κ: 0.353-0.453) for Baumgaertner reduction quality assessment, and fair-moderate (fleiss κ: 0.365-0.456) for Chang reduction quality assessment. Intra-observer reliabilities were found to be moderate to good for TAD and calTAD measurements, moderate for CDA measurements, good to excellent for AP and lateral zone and Baumgaertner reduction quality criteria assessments and good for Chang reduction quality criteria assessment.

DISCUSSION:
There is no large data on the subject of the present study in the literature. There is a single article evaluating the reliability of more than one evaluation criteria in which only two observers' findings were evaluated in the literature. As in our study, more accurate inferences can be made with the increase in the number of observers. Unlike the literature, the inter-observer reliabilities of TAD and calTAD which have a very important place in the postoperative evaluation of these fractures were found as "moderate" instead of "excellent" in our study.

CONCLUSIONS:
It was determined that only the inter-observer reliability of antero-posterior zone evaluation was substantial. The inter-observer reliability of all other measurements and evaluations were fair to moderate.

Klíčová slova: Intertrochanteric fractures, reliability, measurements, reduction quality; TAD, calTAD

Zveřejněno: 15. únor 2022  Zobrazit citaci

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
TURGUT A, FILIBELI M, KUMBARACI M, BILGIN E, HANCIOĞLU S, KALENDERER Ö. Spolehlivost hodnocení na operatérovi závislých faktorů, jež ovlivňují mechanické selhání v léčbě intertrochanterické zlomeniny femuru. Acta Chir Orthop Traumatol Cech. 2022;89(1):75-80. doi: 10.55095/achot2022/010. PubMed PMID: 35247248.
Stáhnout citaci

Reference

  1. Baumgaertner MR, Curtin SL, Lindskog DM. Intramedullary versus extramedullary fixation for the treatment of intertrochanteric hip fractures. Clin Orthop Relat Res. 1998;348:87-94. Přejít k původnímu zdroji...
  2. Baumgaertner MR, Curtin SL, Lindskog DM, Keggi JM. The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip. J Bone Joint Surg Am. 1995;77:1058-1064. Přejít k původnímu zdroji... Přejít na PubMed...
  3. Chang SM, Zhang YQ, Ma Z, Li Q, Dargel J, Eysel P. Fracture reduction with positive medial cortical support: a key element in stability reconstruction for the unstable pertrochanteric hip fractures. Arch Orthop Trauma Surg. 2015;135:811-818. Přejít k původnímu zdroji... Přejít na PubMed...
  4. Cleveland M, Bosworth DM, Thompson FR, Wilson HJ, Ishizuka T. A ten-year analysis of intertrochanteric fractures of the femur. J Bone Joint Surg Am. 1959;41:1399-1408. Přejít k původnímu zdroji... Přejít na PubMed...
  5. De Bruijn K, den Hartog D, Tuinebreijer W, Roukema G. Reliability of predictors for screw cutout in intertrochanteric hip fractures. J Bone Joint Surg Am. 2012;94:1266-1272. Přejít k původnímu zdroji... Přejít na PubMed...
  6. FeldinszkÁ J, Jacko P, Barinka J, Kilian M, Šimko P. Comparison of 2-screw implant and antirotational blade implant in treatment of trochanteric fractures. Acta Chir Orthop Traumatol Cech. 2020;87:268-272. Přejít k původnímu zdroji... Přejít na PubMed...
  7. Kashigar A, Vincent A, Gunton MJ, Backstein D, Safir O, Kuzyk PR. Predictors of failure for cephalomedullary nailing of proximal femoral fractures. Bone Joint J. 2014;96-B:1029-1034. Přejít k původnímu zdroji... Přejít na PubMed...
  8. Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med. 2016;15:155-163. Přejít k původnímu zdroji... Přejít na PubMed...
  9. Kuzyk PR, Zdero R, Shah S, Olsen M, Waddell JP, Schemitsch EH. Femoral head lag screw position for cephalomedullary nails: a biomechanical analysis. J Orthop Trauma. 2012;26:414-421. Přejít k původnímu zdroji... Přejít na PubMed...
  10. Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33:159-174. Přejít k původnímu zdroji... Přejít na PubMed...
  11. Mao W, Ni H, Li L, He Y, Chen X, Tang H, Dong Y. Comparison of Baumgaertner and Chang reduction quality criteria for the assessment of trochanteric fractures. Bone Joint Res. 2019;8:502-508. Přejít k původnímu zdroji... Přejít na PubMed...
  12. Sambandam SN, Chandrasekharan J, Mounasamy V, Mauffrey C. Intertrochanteric fractures: a review of fixation methods. Eur J Orthop Surg Traumatol. 2016;26:339-353. Přejít k původnímu zdroji... Přejít na PubMed...
  13. Turgut A, Kalenderer Ö, Günaydin B, Önvural B, Karapinar L, Ağuş H. Fixation of intertrochanteric femur fractures using Proximal Femoral Nail Antirotation (PFNA) in the lateral decubitus position without a traction table. Acta Orthop Traumatol Turc. 2014;48:513-520. Přejít k původnímu zdroji... Přejít na PubMed...
  14. Turgut A, Kalenderer Ö, Karapinar L, Kumbaraci M, Akkan HA, Ağuş H. Which factor is most important for occurrence of cutout complications in patients treated with proximal femoral nail antirotation? Retrospective analysis of 298 patients. Arch Orthop Trauma Surg. 2016;136:623-630. Přejít k původnímu zdroji... Přejít na PubMed...
  15. Yam M, Chawla A, Kwek E. Rewriting the tip apex distance for the proximal femoral nail anti-rotation. Injury. 2017;48:1843-1847. Přejít k původnímu zdroji... Přejít na PubMed...