Opakovaná sterilizace artikulačních komponent kloubních náhrad etylenoxidem nemá vliv na strukturu a vlastnosti UHMWPE

Repeated Sterilization of Articular Components of Joint Replacements with Ethylene Oxide Does Not Affect the Structure and Properties of UHMWPE

P. FULÍN1, M. ŠLOUF2, V. GAJDOŠOVÁ2, P. NĚMEČEK2, S. KREJČÍKOVÁ2, D. POKORNÝ1
1 l. ortopedická klinika 1. lékařské fakulty Univerzity Karlovy a Fakultní nemocnice v Motole, Praha
2 Ústav makromolekulární chemie Akademie věd České republiky, Praha

ABSTRACT

PURPOSE OF THE STUDY

In clinical practice UHMWPE is the most commonly used material for manufacturing articular components of joint replacements. The purpose of this study is to find out whether repeated ethylene oxide sterilization results in oxidative degradation of UHMWPE or not and also whether the oxidative degradation of various types of ethylene oxide-sterilized UHMWPE depends on storage time or not.

MATERIAL AND METHODS

The set included 12 samples of UHMWPE (three samples with different modifications (virgin PE, with E vitamin and cross-linked with thermal treatment) and different number of sterilizations (0x-3x)). The set also included 8 samples of commercial components of hip or knee replacements sterilized with ethylene oxide and stored for different storage periods. The oxidative degradation was assessed by infrared microspectroscopy, based on which the oxidation index (OI), transvinylene index (VI), crystallinity index (CI) and E vitamin index (EI) were calculated. Mechanical properties of UHMWPE were obtained through microhardness measurements. Statistical processing of the results was performed.

RESULTS

In all the samples, very low oxidative degradation values were reported (most OI values < 0.1). All radiation crosslinked UHMWPE samples showed an increased VI index and a slightly lower crystallinity index. All unmodified samples (irrespective of whether or not and how many times or how long ago the samples were sterilized with EtO) had almost zero value of VI. Changes in crystallinity were negligible (in the rage of 0.56-0.63), which required very accurate measurements of micromechanical properties. Yet, linear correlation was established between microhardness and crystallinity.

DISCUSSION

All the mentioned indices changed as anticipated: OIs were very low and slightly increased with time of storage, VIs of radiation crosslinked samples grew in proportion to the total gama radiation dose, CIs decreased in samples thermally treated by remelting, and EIs were very low due to negligible concentration of stabiliser (0.1%) in the samples of medical grade UHMWPE.

CONCLUSIONS

All samples showed zero or minimum oxidative degradation. This confirmed that neither ethylene oxide sterilization, nor multiple EtO sterilization or longer storage of polymer after ethylene oxide sterilization result in major oxidative degradation.

Key words: UHMWPE, ethylene oxide, sterilization, oxidation, infrared spectroscopy, microhardness

Zpět


Nabídka nakladatelství:

 

Peter Wendsche, Radek Veselý et al.
Traumatologie

Traumatologie

Druhé, přepracované a rozšířené vydání

 

Jan Lebl, Zdeněk Šumník, Ondřej Souček, Hana Malcová, Klára Maratová, Jana Plešková, Štěpánka Průhová, Jan Štulík, Lukáš Wagenknecht
Onemocnění skeletu u dětí

Onemocnění skeletu u dětí

Motolské pediatrické semináře 4

 

Jiří Skála-Rosenbaum, Valér Džupa, Martin Krbec et al.
Zlomeniny proximálního femuru

Zlomeniny proximálního femuru

 

Jiří Kříž et al.
Poranění míchy

Poranění míchy

Příčiny, důsledky, organizace péče