Acta Chir Orthop Traumatol Cech. 2008; 75(3):205-211 | DOI: 10.55095/achot2008/039
Mid-Term Results of Cementless Ultima Cups in Primary Total Hip ReplacementOriginal papers
- Ortopedické oddělení Nemocnice Pelhřimov
PURPOSE OF THE STUDY:
To present the results of primary total hip arthroplasty (THA) with use of the hemispheric threaded Ultima cup at 8.5-year follow-up.
MATERIAL:
Between 1996 and 1999, a total of 40 Ultima acetabular components were used in 33 patients (15 men and 18 women). The average age at the time of surgery was 61.2 years (range, 45 to 71). By the end of 2006, 36 cups were assessed. Indications for the primary THA procedure included primary arthritis in 32 hips, post-traumatic arthritis in three, post-dysplastic arthritis in two, femoral neck fracture in two and rheumatoid arthritis in one.
METHODS:
The average follow-up was 8.5 years (range, 7 to 10). The indication criteria for primary implantation were evaluated, i.e., body mass index, patients' activity and their age. THA outcome was evaluated on the basis of clinical rating (Harris hip score) and pain assessment and on radiographic findings on which the acetabulum before and cup position after THA were compared, and potential changes in cup position, acetabular cup loosening or para-articular ossifications were observed.
RESULTS:
Good outcomes were found in 88.9 % of the hips. In most, cup position was satisfactory. One acetabular component was implanted in a varus position (34°). An intraoperative complication included fracture of the greater trochanter, which was left untreated for spontaneous healing.
The early-postoperative complications were one dislocation, two wounds with serous secretion not requiring revision surgery, and one large haematoma. Stem fracture as a late complication was recorded in two THAs. Radiolucencies were found in six hips and paraarticular ossifications in four. Migration of the cup with protrusion into the acetbulum occured in one patient 3 years after THA. Four cups showed aseptic loosening within 3 to 7 years of surgery. No infection was recorded in this THA group.
DISCUSSION:
Information on mid- and long-term outcomes of the use of cementless acetabular cups in the relevant literature is sparse. The Ultima cup is dealt with only in the report by Pazdírek et al., whose results are in agreement with ours, but involve a lower number of implanted cups and a shorter follow-up. In this study, loosening was recorded only in cups larger than 32 mm in diameter. In comparison with our group of cementless BMT acetabular components, the use of cementless Ultima cups gives better results at mid-term follow-up.
CONCLUSIONS:
At 8.5-year follow-up, 88.9 % of the Ultima acetabular components were without radiographic signs of loosening. Attention should be paid primarily to THA patients with a thin wall acetabulum requiring a larger cup and a 32-mm liner.
Keywords: cementless cup, hemispherical threaded cup, total hip athroplasty
Accepted: January 8, 2008; Published: June 1, 2008 Show citation
References
- BARTONÍČEK, J., DŽUPA, V.: Problematika polyetylenu u totálních náhrad kyčelního kloubu. Acta Chir. orthop. Traum. čech., 66: 350-355, 1999.
- BROOKER, A. F., BOWERMAN, J. W., ROBINSON, R. A., RILEY, L. H.: Ectopic ossification Following Total Hip Replacement - Incidence and Method of Classification. J.Bone Jt Surg., 55-A: 1629-1632, 1973.
Go to original source...
- ČECH, O., PAVLANSKÝ, R.: Aloplastika kyčelního kloubu. Praha, Avicenum 1992.
- ČECH, O., DŽUPA, V., SVATOŠ, F.: Necementovaná jamka endoprotézy kloubu CLS Spotorno - pětileté výsledky. Acta Chir. orthop. Traum. čech, 68: 10-17, 2001.
- ČECH, O., DŽUPA, V.: Revizní operace náhrad kyčelního kloubu. Praha, Galén 2004, 119-136.
- DUNGL, P. a kol.: Ortopedie. Praha, Grada Publishing 2005, 922-925.
- GALLO, J., KAMÍNEK, P., ZAPLETALOVÁ, J., NOVOTNÝ, R., ČECHOVÁ, I., DITMAR, R.: Předčasné selhání polyetylenové vložky u TEP kyčelního kloubu ABG I. Acta Chir. orthop. Traum. čech., 70: 343-349, 2003.
- GALLO, J., KAMÍNEK, P., ZAPLETALOVÁ, J., ČECHOVÁ, I., ŠPIČKA, J., DITMAR, R.: Je osteolýza kolem stabilní TEP kyčelního kloubu asymptomatická? Acta Chir. orthop. Traum. čech., 71: 20-25, 2004.
- GALLO, J., LANDOR, I., VAVŘÍK, P.: Současné možnosti prevence infektu kloubních náhrad. Acta. Chir. orthop. Traum. čech., 73 :229-236, 2006.
Go to original source...
- GRUEN, T. A., McNEICE, G. M., AMSTUTZ, H. C.: "Modes of failure" of Cemented stem Type Femoral Components. Radiographic Analysis of Loosening. Clin. Orthop., 141: 17-27, 1969.
Go to original source...
- HAJNÝ, P., ŠTEDRÝ, V.: Necementovaná náhrada kyčelního kloubu ABG: popis implantátu, střednědobé výsledky. Acta Chir. orthop. Traum. čech., 66: 181-186, 1999.
- HART, R., ROZKYDAL, Z.: Dlouhodobé výsledky totální endoprotézy kyčelního kloubu Poldi. Acta Chir. orthop. Traum. čech., 66: 139-145, 1999.
- HIPP, E., RECHL, H., PLOTZ, W., REEG, S., TRAGER, J.: Current State of Cementless Hip Prostheses. Acta Chir. orthop. Traum. čech., 61: 103-108, 1994.
- JAHODA, D., NYČ, O., POKORNÝ, D., LANDOR, I., SOSNA, A.: Antibiotika v prevenci infekčních komplikací u operací kloubních náhrad. Acta Chir. orthop. Traum. čech., 73: 108-113, 2006.
Go to original source...
- KOPECKÝ, I.: Paraartikulární osifikace totálních endoprotéz kyčelního kloubu. I. část: etiologie a prevence. Acta Chir. orthop. Traum. čech., 61: 224-226, 1994.
- KRBEC, M., ČECH, O., DŽUPA, V.: Reoperace cementované totální endoprotézy kyčle I. RTG - diagnostika a klasifikace uvolnění. Acta Chir. orthop. Traum. čech., 59: 23-26, 1992.
- KRBEC, M., ADLER, J., MESSNER, P., jr.: Kostní štěpy v revizní endoprotetice kyčle. Acta Chir. orthop. Traum. čech., 70: 83-88, 2003.
- KRBEC, M., ČECH, O., DŽUPA, V., PACOVSKÝ, V., KLÉZL, Z.: Infekční komplikace TEP kyčelního kloubu. Acta Chir. orthop. Traum. čech., 71: 179-188, 2004.
- KUBÁT, P.: Šestileté výsledky implantace závitořezné jamky Walter-Motorlet. In: 4. národní kongres ČSOT, Hradec Králové 2000. Souhrny přednášek 53.
- LANDOR, I., VAVŘÍK, P., SOSNA, A., BEK, J.: Totální náhrada kyčelního kloubu typu Walter - dlouhodobé výsledky. Acta Chir. orthop. Traum. čech., 68: 281-288, 2001.
- LETOCHA, J., BĚHOUNEK, J., SKOTÁK, M..: Zkušenosti s použitím necementované jamky Beznoska typu "ježek" na ortopedickém oddělení Pelhřimov. Acta Chir. orthop. Traum. čech., 74: 195-201, 2007.
Go to original source...
- MORSCHER, E. W.: Noncemented Acetabular Fixation in Primary Total Hip Replacement. In: Morscher, E. W. (ed.): Endoprosthetics. Berlin, Springer 1995, 143-179.
Go to original source...
- NUNN, D., FREEMANN, M., HILL, P., EVANS, S.: The Measurement of Migration of the Acetabular Component of Hip Prosthesis. J. Bone Jt Surg., 71-B: 629-631, 1989.
Go to original source...
Go to PubMed...
- PAZDÍREK, P., BARTONÍČEK, J., DŽUPA, V.: Necementovaná závitořezná jamka Ultima: zkušenosti prvních pěti let. Acta Chir. orthop. Traum. čech., 68: 24-30, 2001.
- SOSNA, A., RADONSKÝ, T., POKORNÝ, D., VEIGL, D., HORÁK, Z., JAHODA, D.: Polyetylénová choroba. Acta Chir. orthop. Traum. čech., 70: 6-16, 2003.
- SOSNA, A., ČECH, O.: Operační přístupy ke skeletu pohybového aparátu. Praha, Avicenum 1987, 110-115.
- ŠPIČKA, J., GALLO, J., ČECHOVÁ, I., KAMÍNEK, P., MIKULÍK, J.: Operační léčba osteolýzy kolem stabilní necementované TEP kyčelního kloubu. Acta Chir. orthop. Traum. čech., 72: 228-233, 2005.
- ŠTEDRÝ, V., HAJNÝ, P.:Dlouhodobé výsledky TEP kyčelního kloubu POLDI. Acta Chir. orthop. Traum. čech., 64: 282-291, 1997.
- TRČ, T.: První zkušenosti s použitím necementované endoprotézy kyčle s porézním povrchem. Acta Chir. orthop. Traum. čech., 64: 144-153, 1997.