Acta Chir Orthop Traumatol Cech. 2008; 75(4):282-287 | DOI: 10.55095/achot2008/050

První zkušenosti s totální náhradou zápěstí naší konstrukcePůvodní práce

J. PECH*, D. VEIGL, J. DOBIÁŠ, S. POPELKA, V. BARTÁK
I. ortopedická klinika 1. LF UK a FN Motol, Praha

PURPOSE OF THE STUDY:
To present the results of total wrist replacement with a cementless prosthesis designed by us.

MATERIAL AND METHODS:
A group of five men and 27 women treated between 2004 and 2007 was evaluated. The average age was 51.6 years and follow-up ranged from 4 to 38 months, with an average of 9.4 months. Indications for surgery included wrist destruction due to rheumatoid arthritis, arthritis or psoriatic arthropathy.
Subjective evaluation recorded satisfaction of the patients with the wrist range of motion, cosmetic appearance of the hand and pain relief. The range of motion was assessed with a goniometer at 4 months of follow-up on average. Radiographs were made immediately after surgery and then at 6 and 12 months post-operatively.

RESULTS:
All patients reported pain relief, 23 were satisfied with wrist mobility and hand appearance, and seven were only partly satisfied and would have preferred a greater range of motion. Two patients were dissatisfied with an ulnar deviation of the carpal axis and therefore revision arthroplasty of the radial ligamentary and tendinous structures was planned. No aseptic or septic loosening of the implant or its migration was found.

DISCUSSION:
The current total wrist arthroplasty has evolved from Swanson silastic implants to prostheses with a metal-on-polyethylene bearing whose active surfaces are firmly fixed in bone. In today's Europe, the most frequently used prostheses are Meuli implants or the latest modification of the Guepar implant, which is very close to the Universal 2 implant. In the USA since 1985, Professor's Beckenbaugh's BIAX implant has continuously been improved. The implant designed by us is based on all well-tired implant components and aims at being as similar to the anatomical wrist structures as possible. The socket is fixed in the distal radius. The carpal components involve the main fixation shaft for the third metacarpal and a short antirotation pin for the second metacarpal base. A metal head is assembled onto a tapered shaft extending from the carpal component transversal zone. The implant is made of titanium alloy, in some parts coated with hydroxyapatite; the sockets has a polyethylene liner.

CONCLUSION:
Total wrist replacement is usually preceded, particularly in surgery for rheumatoid arthritis, by operations intended to prevent or slow down gradual wrist destruction. The usual salvage procedures include surgery on soft tissues (synovectomy, tenodesis, tendon reconstruction) and/or bone (limited wrist arthrodesis, Sauve-Kapandji procedure). These procedures usually have a temporary effect and total replacement will be the next step in surgical treatment. For the most severe destruction associated with subluxation and wrist instability, total arthrodesis in a neutral wrist position still remains the method of choice. The first results with use of the implant of our design show that this total wrist replacement allows us to preserve or restore the wrist range of motion and to improved grip strength.

Klíčová slova: total wrist arthroplasty, wrist surgery, prosthesis design, rheumatoid arthritis

Přijato: 11. únor 2008; Zveřejněno: 1. srpen 2008  Zobrazit citaci

ACS AIP APA ASA Harvard Chicago Chicago Notes IEEE ISO690 MLA NLM Turabian Vancouver
PECH J, VEIGL D, DOBIÁŠ J, POPELKA S, BARTÁK V. První zkušenosti s totální náhradou zápěstí naší konstrukce. Acta Chir Orthop Traumatol Cech. 2008;75(4):282-287. doi: 10.55095/achot2008/050. PubMed PMID: 18760084.
Stáhnout citaci

Reference

  1. ALNOT, Y. et al.: L'arthroplastie totale GUEPAR de poignet dans la polyarthrite rheumatoide. Rev. Chir. Orthop., 74: Supl. 2, 340-345.
  2. ALNOT, Y.: GUEPAR total wrist arthroplasty in rheumatoid arthritis. In: Simmen, B.R., Hagena, F. (eds.): The writ in rheumatoid arthritis. Rheumatology, 17. Basel, Karger 1992, 205-213.
  3. BECKENBAUGH, R. D.: Implant arthroplasty in the reumatoid hand and wrist; Current state of the art in the United States. J. Hand Surg., 8-A: 675-678, 1983. Přejít k původnímu zdroji... Přejít na PubMed...
  4. BOROVANSKÝ, L. et al.: Soustavná anatomie člověka. 3. vyd. Praha, Zdrav. nakl. 1967, 153-157.
  5. COBB, T., BECKENBAUGH, R.: Biaxial total wrist arthroplasty. J. Hand Surg., 21-A: 1011-1021, 1996. Přejít k původnímu zdroji... Přejít na PubMed...
  6. DIVELBISS, B., SOLLERMAN, CH., ADAMS, B.: Early results of the Universal total wrist arthroplasty in rheumatoid arthritis. J. Hand Surg., 27-A: 195-204, 2002. Přejít k původnímu zdroji... Přejít na PubMed...
  7. FATTI, J. F. et al.: Long - term results of Swanson interpol. Wrist arthroplasty. J. Hand Surg., 16-A: 432-437, 1991. Přejít k původnímu zdroji... Přejít na PubMed...
  8. FOURASTIER, J., LE BRETON, L., LANGLAIS, F. et al.: Guepars total radiocarpal prosthesis in the surgery of the rheumatoid wrist. Rev. Chir. Orthop. Repat. Appar Mot., 82: 108-115, 1996.
  9. KISTLER, U., WEISS, A., SIMMEN, B., HERREN, D.: Long-term results of silicone wrist arthroplasty in patiens with rheumatoid arthritis. J. Hand. Surg., 30-A: 1282-1287, 2005. Přejít k původnímu zdroji... Přejít na PubMed...
  10. LIRETTE, R., KINNARD, P.: Biaxial total wrist arthroplasty in rheumatoid arthritis. J. Hand Surg, 38-A: 51-53, 1995.
  11. LUTONSKÝ, M., PELLAR, D.: Artrodéza karpometakarpálního kloubu palce ruky. Acta Chir. orthop. Traum. čech., 73: 345-349, 2006. Přejít k původnímu zdroji...
  12. MEULI, H. CH.: Meuli total wrist arthroplasty. Clin. Orthop., 149: 118-125, 1984. Přejít k původnímu zdroji...
  13. MEULI. CH., FERNANDEZ, D. L.: Uncemented total wrist arthroplasty. J. Hand Surg., 20-A: 115-122, 1995. Přejít k původnímu zdroji... Přejít na PubMed...
  14. PECH, J., POPELKA, S. et al.: Revmatochirurgie ruky. Praha, Schola Nova Comenium 2000.
  15. DOBIÁŠ, J., PECH, J., POPELKA, S.: Výsledky implantace silastikových náhrad MCP kloubů u revmatiků. Acta Chir. orthop. Traum. Čech., 74: 278-86, 2007. Přejít k původnímu zdroji...
  16. PECH, J., POPELKA, S., SOSNA, A., VAVŘÍK, P.: První zkušenosti se Swansonovou silastikovou náhradou zápěstního kloubu. Acta Chir. orthop. Traum. čech., 62: 99-105, 1995.
  17. POPELKA, S., RYBKA, V.: Revmatochirurgie, Praha, Avicenum, 1989.
  18. RIZZO, M., BECKENBAUGH, R.: Results of biaxial total wrist arthroplasty with a modified metacarpal stem. J. Hand Surg., 28-A: 577-584, 2003. Přejít k původnímu zdroji... Přejít na PubMed...
  19. SWANSON, A. B.: Flexible implant resection arthroplasty. The Hand, 4: 119-132, 1972. Přejít k původnímu zdroji... Přejít na PubMed...
  20. SWANSON, A. B. et al.: Flexible implant arthroplasty of the radiocarpal joint. In: Simmen, B. R., Hagena, F. (eds.): The wrist in rheumatoid arthritis. Rheumatology, 17. Basel, Karger 1992, 151-161.
  21. SWANSON, A. B., SWANSON DE GROOT, G., MAUPIN, B. K.: Flexible implant arthroplasty of the radiocarpal joint. Clin. Orthop. Rel. Res., 187: 94-99, 1984. Přejít k původnímu zdroji...